Troubleshooting
Opening the lid without power
Error messages
In the event of power failure, the vacuum
system vents automatically, the evaporation
vessel may take up to two minutes
to stop
spinning.
When the evaporation vessel is stationary,
press down on the lid and insert a narrow
tool into the hole near the lid latch
mechanism and to release the lock by
pushing the latch. Withdraw the tool then lift
the lid.
The system notifies the user of any errors
that occur whilst the evaporator is operating
by flashing the blue condenser illumination.
If the Rocket 4D detects a safety critical
error, it will stop safely and show details of
the problem on the screen. If a non-critical
error occurs, it will be flagged but the
evaporation process will continue
uninterrupted.
In all cases, messages are displayed on the
screen after the evaporator stops. A log of
the last ten errors can be viewed via the
Maintenance
Menu and an error log is
included in the
System Info
which can be
downloaded via the
File Transfer
menu.