User Manual
Rocket 4D Evaporation System
04-7848 Issue 2-4 August 2014
21
Options menu
View last method graph
Go to the
Method Select
menu and select
Options
. The menu offers the following
settings that affect the way information is
presented on the Screen.
Go to the
Method Select
menu and select
View Last Method Graph
. The Screen
shows the graph (∆T against time) for the
last completed method, refer to
Condenser
,
DeltaT
.
About menu
Go to the
Method Select
menu and select
About
. The evaporator type, serial number,
software versions and contact information
are shown.
Options
Parameter
Setting
Effect
Power Save
On / Off
Switches off power after 30 minutes if no user intervention
Key Beep
On / Off
Disables key beep
Audible Alarms
On / Off
Disables audible alarms
Pre Run Checks
On / Off
Displays shows safety check list before allowing run to start
Magnified Display
On / Off
Graphic / numeric display replaced by timer count up (or count
down if end time is defined) after five minutes with no
intervention from user
Commissioning
On / Off
Displays information useful during installation. The option to
disable this function is offered at subsequent start-ups
Sys Use Checks
On / Off
Displays extended check list at start of every 10
th
run
Pressure Units
M / t
Displays pressure in millibar or Torr
Temperature Units
C / F
Displays temperature in centigrade or Fahrenheit
Waste full detect
On / Off
Enables operation of optional waste full detector
Inert Gas Blanket
N/A
Not applicable for Rocket 4D
Inert Gas Purge
N/A
Not applicable for Rocket 4D
Change IGP
Password
N/A
Not applicable for Rocket 4D