Rocket 4D Evaporation System
User Manual
32
04-7848 Issue 2-4 August 2014
Feed system
Hoses
Check the product solution and rinse solvent
feed hoses are undamaged (no sharp bends,
no kinks or constrictions that could affect
fluid flow).
Make sure the hose weights are attached
and positioned so the hoses sink to the
bottom of the product solution and rinse
solvent containers. Make sure it is not
possible for the ends of the hoses to form
seals against the side or bottom of the
reservoir due to suction.
Feed sensor
Using a soft brush, remove any dust from the
optical liquid feed sensor which is attached to
the rotary joint feed hose.
If the feed hose becomes discoloured or is
replaced, the feed sensor must be
recalibrated. Before recalibrating, make sure
the hose is dry and is fully engaged in the
sensor. Using the select control, go to the
Maintenance
menu and follow the on screen
prompts to calibrate the feed sensor.
For descriptions of hose fittings, see
Hose
connection method
in
Unpacking and
Setting-up
.
Software upgrade
Software upgrades will be sent by Email.
Contact your local Genevac representative
for further information.
To upload software:
1.
Using a PC, copy the software
upgrade files to a root folder called
\ROCKET\
on a USB data key.
2.
Switch off the evaporator.
3.
Insert the datakey into the evaporator
USB socket (under the left side of the
control panel).
USB data key
4.
Switch on the evaporator, when the
system has validated the upgrade,
press the start button.
5.
Remove the datakey.
Optical feed