Rocket 4D Evaporation System
User Manual
14
04-7848 Issue 2-4 August 2014
Draining the condenser
In normal operation, draining is fully automatic and requires no user intervention. However,
if manual draining is required, go to the
Method Select
menu, select the
Drain Condenser
option then press the start button
.
Observe the condenser to make sure all solvent has drained, then press any button to close
the drain valve (after six minutes the condenser drain valve closes automatically).
Make sure the waste solvent container is empty before starting the evaporator. Do not
disconnect the waste solvent container while the evaporator is operating as the drain valve
may open without warning.
Pouring stand
For easy pouring of concentrated product from the evaporation vessel rotor, place on the
pouring stand with the spout positioned as shown. Lift the front of the stand and remove
the stopper from the spout, then lift the back of the stand to pour.
Evaporation vessel (bowl rotor) with pouring stand
Warning:
Entrapment hazard, keep hands away from the sides of the pouring
stand when raising or lowering.
Drain Condenser
Drain Will Commence On
Next Key P ress…
P lease Ensure Waste Connected
to Start,
to Return