Note:
For random or waveform variation functions, the maximum PDL setting must be
greater than the minimum PDL setting. To satisfy this requirement, if a “set maximum”
command such as *SIN:MXA f.f# is used to set the maximum PDL to a value lower than
the current minimum PDL setting, the minimum PDL setting will then be changed to
Min = Max
−
0.1 dB. Similarly, if a “set minimum” command such as *SIN:MNA f.f# is
used to set the minimum PDL to a value greater than the current maximum PDL setting,
the maximum PDL setting will then be changed to Max =Min +0.1 dB.
Error Code Definitions
E00 No error (State OK)
E01 Undefined
command
E05 Value out of range
RS-232 command notes:
1.
RS-232 port uses asynchronous framing, 8 data bits, no parity bit, 1 stop bit and
baud rate: 9600 bps.
2.
Only one command is allowed in each command string.
3.
The PDLE-101 should respond to control commands with one of the error codes
listed above. If a command is a status request, the computer will return either the
requested information or an error code of form *Enn#; see the error definition list
above.
4.
While the PDLE-101 is processing a command and sending back a response
ASCII code, any additional commands sent by the computer will be ignored.
4.4.1.3
RS-232 troubleshooting
If a problem occurs during RS-232 control, please check the following:
1.
Cable should be straight-wired: Select a straight-wired (pin-to-to pin wired) cable;
2.
Verify active Comm Port (COM1, COM2, etc.);
3.
Verify Comm Port settings: 8 data bits, 1 stop bit, no parity bits;
4.
Check baud rate: must be 9600 bps;
5.
Check command format: a correct command should begin with “*” and end with
“#” or “?”.
4.4.2 USB control
4.4.2.1 Procedure for Installing USB VCP driver
1.
Copy VCP driver from CD-ROM to the control computer.
2.
Turn on instrument power supply and plug in USB cable; this should bring up the
“Found New Hardware Wizard”:
Document #: GP-UM-PDLE-101-11
Page 30 of 61
Содержание PDLE-101
Страница 31: ...3 Click Next 4 Click Next Document GP UM PDLE 101 11 Page 31 of 61 ...
Страница 32: ...5 Check Specify a location then click Next Document GP UM PDLE 101 11 Page 32 of 61 ...
Страница 36: ...12 Click Next 13 Click Finish Document GP UM PDLE 101 11 Page 36 of 61 ...
Страница 38: ...In this example the USB device is connected to COM3 Document GP UM PDLE 101 11 Page 38 of 61 ...
Страница 47: ... Properties on General tab Internet Protocol Properties Document GP UM PDLE 101 11 Page 47 of 61 ...
Страница 49: ... Properties on General tab Document GP UM PDLE 101 11 Page 49 of 61 ...