Section 4. Operation Instructions:
Warning:
•
Never look into the light source fiber connector when the light source is
turned on. THE OUTPUT LIGHT FROM A HIGH POWER LASER IS
HARMFUL TO HUMAN EYES. Please follow industry standard
procedures when operating a high power laser source.
•
The PDLE-101 is designed for indoor use only. Avoid water condensation or
liquid spills during storage and operation.
4.1 Unpacking
Inspect PDLE-101 for any physical damage due to shipping and transportation. Contact
carrier if any damage is found. Check the packing list to see if any parts or accessories
are missing.
4.2 Front Panel Operation
General:
1.
Make sure local AC voltage matches the AC voltage requirement of the PDLE-
101 system. If not, do not proceed. Contact General Photonics immediately.
The standard model uses 100-240 VAC, 50/60Hz, 1A Fuse.
2.
Connect the power cord and plug it into the wall receptacle. Make sure the
ground pin of the power cord is connected to earth ground.
3.
On power-up, the initialization screen, which shows the hardware and firmware
versions, will display for a few seconds before the main display screen appears.
→
GENERAL PHOTONICS
PDLE-101 VER 1.1
PDL : 12.3 dB
λ
: 1550 nm
Initialization
Screen
Main
Display
Screen
The main display screen shows the current PDL and wavelength settings.
4.
Connect the input and output fibers to the PDLE-101. It is important to clean the
fiber connectors using industry standard procedures before connecting them. In
the case of a high power laser source, turn off the optical power source before
connector cleaning. The PDLE-101 is a unidirectional device. Make sure that the
light source is connected to the input port.
Document #: GP-UM-PDLE-101-11
Page 12 of 61
Содержание PDLE-101
Страница 31: ...3 Click Next 4 Click Next Document GP UM PDLE 101 11 Page 31 of 61 ...
Страница 32: ...5 Check Specify a location then click Next Document GP UM PDLE 101 11 Page 32 of 61 ...
Страница 36: ...12 Click Next 13 Click Finish Document GP UM PDLE 101 11 Page 36 of 61 ...
Страница 38: ...In this example the USB device is connected to COM3 Document GP UM PDLE 101 11 Page 38 of 61 ...
Страница 47: ... Properties on General tab Internet Protocol Properties Document GP UM PDLE 101 11 Page 47 of 61 ...
Страница 49: ... Properties on General tab Document GP UM PDLE 101 11 Page 49 of 61 ...