TS4000
33
NOTE:
When the Sensor Range is changed, the Warning and Alarm set points are
automatically scaled to the new range. The unit must now be calibrated to the new
range. Refer to Section 4.9.
CAUTION:
If the user does not complete the menu cycle a fault occurs. For more information
about faults and fault codes, refer to Section 7.1.
4.6.2 Warning Relay Settings
The user can adjust the Warning Relay Settings for all sensors. The factory default settings and
adjustment range for all gases other than O
2
are as follows:
•
Non-latching (default)
•
De-energized (default)
•
30% FS set point (default)
•
5% of FS (minimum)
•
Alarm relay set point (maximum)
The default settings and adjustment range for O
2
are:
•
19.5% by volume (default)
•
Alarm Relay set point (minimum)
NOTE:
In general, most configuration procedures apply to all sensors. However, there are
instances when a unique procedure is required for a given sensor or sensors.
Wherever applicable, both generic and unique procedures are detailed in this manual.