8. Remove the gas by closing the valve on the cylinder and remove the cup allowing the
sensor to read clean air. The display changes from “
CC
” to indicate a few ppm and
then drops to “
0
”.
The unit is now calibrated and the new calibration constants are stored in the EEPROM.
For calibration accessories from General Monitors, refer to Sections 9.5.4 and 9.7.