M02 Digital Indicator
11
In the interface of highest point calibration,press
to enter into analog
display value interface, the display will be 5 digits (initial value is
24000,
means
24.000mA
),press
to input the value measured by the multimeter.
Note: Analog output calibration, highest point must be calibrated.
2.4.2 Serial Interface RS485 Output
Serial Interface RS485 output is optional, please refer to chapter 6.0 for
communication protocol.
RS485 serial interface connection:
This interface is the highest
point calibration of analog
output.
Содержание M02