M02 Digital Indicator
10
and
Current calibration interface
In this interface
,
the
multimeter measured
should be 4.000mA.
Measured
4.000mA
If not,press
If yes,press
In this interface
,
the
multimeter measured
should be 4.000mA.
This
interface
is
current
DA
D/A,press
to adjust the value,
and the current value measured will
vary accordingly.To adjust it when the
measured value to be 4.000mA, then
press
to save and exit.
Then press
to calibrate
12.000mA.
Finish the calibration referring to the process of
4.000mA calibration, and the same for the
calibration of 20mA.