M02 Digital Indicator
55
40023
0022
Zero calibration with weights: write in 1 and calibrate zero
with the current weight.
Read: Absolute millivolt of current load cell
40024
0023
40025
0024
Zero calibration without weights: Write millivolt value at
zero
;
Write in range
(
load cell
3mV/V
:millivolt value range
within
0.02-12.000mV)
millivolt value at zero when read.
40026
0025
40027
0026
Gain calibration with weights. Write weight value(≤max.
capacity). Return millvolt at present weight value when
read
40028
0027
40029
0028
Gain calibration without weights
;
input gain millivolt(load
cell
3mV/V:0.000
<
millivolt<
15.000mV
-
zero
millivolt).Read: millivolt value for gain calibration.
40030
0029
40031
0030
Gain calibration weight without weights. input gain
millivolt(≤max. capacity) Read: weight value for gain
calibration.
40032
0031
40033~40040
0032~0039 reserved
40041
0040
Set point
1 stable or not
(
0
:
no
;
1
:
yes
)
40042
0041
Set point
1
min. duration time
(
0-999 : 0-99.9sec.
)
40043
0042
Set point
1 valid condition
40044~40045
0043~0044
Set point
1
set value 1