M02 Digital Indicator
14
3.2 Flow Chart of Calibration
Normal Status
Password input
Decimal point
2.
After password is input, the indicator will
display
CALON
for one second, then go to
next step.
3.
Press
to select a desired value for
decimal point among
0,
0.0, 0.00, 0.000
and
0.0000, and then
press
to save it and enter next step.
If there’s no need to change the value,
press
directly to enter next step.
1
. Under this status, press
(twice),
indicator
will
display
CAL
,
then
press
to enter password input.