Probes
Venue 50
–
Basic User Manual
2-37
Direction 5446729-100 English
Rev. 7
Connecting the Probe
Probes can be connected whether the console is powered on or
off. To ensure that the ports are not active, place the system in
the image freeze condition.
To connect a probe:
1. Place the probe's carrying case on a stable surface and
open the case.
2. Ensure
the
system is in freeze mode or powered off.
Carefully remove the probe and unwrap the probe cord.
3. DO NOT allow the probe head to hang free. Impact to the
probe head could result in irreparable damage.
Inspect the probe before and after each use for damage or
degradation to the housing, strain relief, lens, seal and
connector. DO NOT use a transducer which appears
damaged until functional and safe performance is verified. A
thorough inspection should be conducted during the
cleaning process.
4. Align the connector with the probe port and carefully push
into place.
Figure 2-33. Connect probe to Venue 50