Customizing Your System
4-28
Venue 50
–
Basic User Manual
Direction 5446729-100 English
Rev. 7
Note: Only one or two bytes can be input in the modality cell. Note: The default value of date range is 0, and
the maximum value is 60. Set date range to 0, the Scheduled Procedure Step Start Date will include today
only. Set the date range to X (an integer number between 0 and 60), the Scheduled Procedure Step Start
Date will include X days prior and from today.
Table 4-20: DICOM Image Storage settings
Preset Parameter
Description
IP Address
Type the IP Address of the DICOM image server.
AE Title
Type the calling AE Title of the DICOM image server.
Port
Type the port of the DICOM image server.
Enable MultiFrame
DICOM
Select to enable multiframe DICOM.
Table 4-21: DICOM Local AE Title
Preset Parameter
Description
DICOM local AE Title
Type the local DICOM called AE Title.
Table 4-19: DICOM Search Criteria settings
Preset Parameter
Description