User Maintenance
6-10
Venue 50
–
Basic User Manual
Direction 5446729-100 English
Rev. 7
Doppler Sensitivity
Measurements were obtained with each transducer over a
range of fluid velocities as indicated in the following table. The
maximum detectable depth of motion was measured and
converted to round-trip sensitivity in dB corresponding to the
depth and transmit frequency. The chart below summarizes the
sensitivities for all transducers in penetration-on mode.
Table 6-11: Doppler Sensitivity Data Summary
Probe
Velocity
Range (cm/s)
Max Depth (cm)
Frequency (MHz)
Sensitivity (dB)
12L-SC
L
15
7
5.00
35
M
63
7
5.00
35
H
110
7
5.00
35
3S-SC
L
15
15.5
2.20
34.1
M
63
15.5
2.20 34.1
H
110
15.5
2.20 34.1
4C-SC
L
15
16.2
2.20
35.64
M
63
16.2
2.20
35.64
H
110
16.2
2.20
35.64
L8-18i-SC
L
15
4.00
6.67
26.68
M
63
4.00
6.67
26.68
H
110
4.00
6.67
26.68
E8CS-SC
L
15
7
4.00
28
M
63
7
4.00
28
H
110
7
4.00
28
10C-SC
L
15
7.5
4.00
30
M
63
7.4
4.00
29.6
H
110
7.4
4.00
29.6
Содержание Venue 50
Страница 4: ...i 2 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7 This page intentionally left blank...
Страница 8: ...i 6 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 14: ...i 12 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 56: ...Introduction Safety 1 42 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 236: ...Customizing Your System 4 34 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 280: ...Probes and Biopsy 5 44 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 327: ......