User Maintenance
6-12
Venue 50
–
Basic User Manual
Direction 5446729-100 English
Rev. 7
We have worked diligently to develop a combination of the
safety measures above and the native security advantage of
Venue 50 to provide a degree of safety against Viruses, Worms,
Trojan Horses, etc., especially for a system used in a
professional hospital grade networking environment that also
typically features its own sufficient safety measures.
Finally, a few points as to why we do not use Anti-Virus software.
The main reasons for not doing so:
•
Customized OS of Venue 50 is natively immune to most
viruses. Only few viruses can run on Venue 50 system.
•
Every Virus scanner is constantly active in the background.
Due to the software-intensive operating system of the
Ultrasound scanner, all computing resources are required
for normal operation of this device. Anti-Virus software
activities would have a negative impact on the system
performance.
•
The operating software of a medical Ultrasound system is
part of an FDA-cleared medical device that requires a
specific release process. Any update of the Anti-Virus
software would mean a change of the system software.
Such change would require an extensive release and
validation process to help ensure that the Anti-Virus update
does not have any impact on the system software
performance and stability.
Содержание Venue 50
Страница 4: ...i 2 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7 This page intentionally left blank...
Страница 8: ...i 6 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 14: ...i 12 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 56: ...Introduction Safety 1 42 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 236: ...Customizing Your System 4 34 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 280: ...Probes and Biopsy 5 44 Venue 50 Basic User Manual Direction 5446729 100 English Rev 7...
Страница 327: ......