Performing an Exam
3-34
Venue 50
–
Basic User Manual
Direction 5446729-100 English
Rev. 7
Crown Rump Length (CRL)
To measure crown rump length, make one distance
measurement:
1. Press
Measure
, then select
CRL
.
Figure 3-29. CRL Screen
2. Touch the screen, and two active calipers display.
3. Move the calipers to the desired position.
4. Press
Set
.
The system displays the value in the Results Window.
Figure 3-30. CRL Measurement