Chapter 2
Overview of the ColorTouch Staff Console and Annunciator
18
Telligence Patient-Staff Communication System Operation Manual
• Code
pink
This setting is configurable, however. Emergency annunciators can
be set to display any call priority.
Coverage modes
Staff consoles and annunciators are further configured with settings
that determine the type of coverage they will provide:
primary
,
annunciate
,
supervision
, and
failure
.
Primary versus annunciate coverage:
Each staff console or
annunciator is set for either
primary
or
annunciate
mode.
• The
primary
coverage device serves as the primary call center
for the duty areas assigned to it. Each network must have at
least one
primary
device to which all nurse calls are directed.
• Devices configured for annunciate mode provide backup
coverage for the area to which they are assigned. This allows
staff members to view and answer calls from locations other
than the nursing station.
Supervision and failure coverage:
S
upervision
and
failure
coverage
modes are used for monitoring critical nurse call equipment. Either
setting can be applied to devices configured either as primary or as
annunciate.
• In supervision mode, the device monitors the communication
function of
staff consoles
or
annunciators
. If a device loses
communication, an error message is sent to the staff console.
Devices operating solely in supervisory mode do not annunciate
patient calls. Instead they signal the failure of the staff consoles
and annunciators assigned to them. Devices can be set to
provide supervision in addition to displaying nurse calls.
• In failure mode, the device monitors patient stations, staff
stations, and peripheral devices for power or data signal failure.
Devices operating solely in failure mode do not annunciate
patient calls. Instead they signal the failure of the devices
assigned to them. This mode can be assigned in addition to
either primary or annunciate, however, which makes it possible
to display nurse calls and device failure notifications on the
same device.
To avoid sending error codes for momentary interruptions of
service, a delay can be applied to supervision and failure mode. By
default, the system waits for 1 minute before sending an error code
to the staff console. This delay is configurable. It can be removed or
set as high as 300 minutes. See the installation manual provided
with your Telligence system
3101010 R1.0 Telligence Patient-Staff
Communications System Installation Manual.
Содержание Telligence Patient-Staff Communications System
Страница 6: ...iv Telligence Patient Staff Communication System Operation Manual ...
Страница 14: ...Chapter 1 General information 8 Telligence Patient Staff Communication System Operation Manual ...
Страница 94: ...Chapter 4 Patient stations 88 Telligence Patient Staff Communication System Operation Manual ...
Страница 122: ...Chapter 5 Staff duty stations 116 Telligence Patient Staff Communication System Operation Manual ...
Страница 136: ...Telligence Patient Staff Communication System Operation Manual 130 ...