![GE Telligence Patient-Staff Communications System Скачать руководство пользователя страница 103](http://html.mh-extra.com/html/ge/telligence-patient-staff-communications-system/telligence-patient-staff-communications-system_operation-manual_132154103.webp)
Chapter 5
Staff/duty stations
Telligence Patient-Staff Communication System Operation Manual
97
• Associated corridor and zone lights show a steady white.
• Associated zone lights emit a slow repeating buzz.
• Staff normal tones sound at staff consoles and annunciators,
and at stations where staff follow is active.
• A staff normal nurse call row appears on the Call screen on staff
consoles and unit annunciators.
Canceling a call from the smart staff/duty station
Press the Cancel button on the staff/duty station to cancel a code
blue alert, staff emergency alert, or a normal priority call placed
from the staff/duty station.
— or —
Press any linked Cancel button to cancel calls from the staff/duty
station.
Pressing Cancel on or for the staff/duty station causes the following:
• Dome lights and staff/duty station lights extinguish.
• Tones sounding at staff consoles/annunciators and staff follow-
activated stations cease.
• The Call screen nurse call row and any corresponding
Reminders screen reminder row on staff consoles and
annunciators disappear.
Notes
• Normal priority calls, including staff normal nurse calls, are
canceled after staff members answer them at staff consoles or
annunciators.
• Pressing the Cancel button for 8 seconds (and 4 feedback tones)
on a staff/duty station or associated room station clears all calls
made from the room, and all reminders for those calls.
Activating staff follow at a staff/duty station
In staff/duty stations configured with an active staff follow mode,
duty mode can be switched on or off. This is not try of permanent
duty stations, in which staff follow is always active.
To turn duty mode on or off:
1. Press the Duty button on the staff/duty station. If any calls are
active in the station’s coverage area, the staff follow tone
sounds and the Status light flashes red to annunciate the
highest priority call.
2. Press the Duty button again to cancel duty mode.
Notes:
Содержание Telligence Patient-Staff Communications System
Страница 6: ...iv Telligence Patient Staff Communication System Operation Manual ...
Страница 14: ...Chapter 1 General information 8 Telligence Patient Staff Communication System Operation Manual ...
Страница 94: ...Chapter 4 Patient stations 88 Telligence Patient Staff Communication System Operation Manual ...
Страница 122: ...Chapter 5 Staff duty stations 116 Telligence Patient Staff Communication System Operation Manual ...
Страница 136: ...Telligence Patient Staff Communication System Operation Manual 130 ...