Assembly and Setup
40
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
2097937-002 Rev G
11 March 2019
If the measurements are significantly out of range, check
for breaks in the power cord or in the internal connections
within the units.
Conducting Leakage Tests
The leakage tests are safety tests to ensure that the equipment
poses no electrical health hazards. Use the table below to
determine which tests apply to the unit under test and the
maximum allowable leakage currents. For international leakage
limits, refer to the internal standards agencies of that particular
country.
If the unit under test fails the leakage tests, do not allow the
customer to use the equipment. Call Tech Support for assistance.
GE Healthcare recommends that you perform these tests:
•
Before applying power for the first time
•
Whenever internal assemblies are serviced
NOTE:
The accuracy of the leakage tests depends on a properly-
wired wall outlet. Do not proceed until you verify the
integrity of the power source.