Calibration
2097937-002 Rev G
T2100-ST1 Treadmill, 110V / T2100-ST2 Treadmill, 220V
167
11 March 2019
Software Interface Instruction T2100-ST Series Calibration
Tools Required
o
Calibration Program (T2100-ST Series Calibration Software)
o
Chalk or some other temporary marking device
o
Computer or laptop (PC)
o
Serial or USB Communication Cable
NOTE:
This task needs to be performed prior to speed or elevation
calibration.
1.
Connect the treadmill to AC power.
2.
Connect the treadmill and PC.
o
If using USB cable, connect a USB cable (GE part
number 2044095-001) between the PC and the T2100-
ST series treadmill and turn on the treadmill.
o
If using RS232 serial cable:
a.
Connect one end of the RS232 serial cable (GE part
number 2097829-109) to the RS232 COM port located
on the lower front panel of the T2100-ST series
treadmill.
b.
Connect the other end of the RS232 serial cable (GE
part number 2097829-109) to a COM port on the PC.
c.
Turn on the Treadmill.
3.
Open T2100-ST Series Calibration Software
by browsing for the
file named
T2100-
ST_Series_Calibration_Software.application
and then double
clicking on it.
NOTE:
This file should be found in the directory in which the
file ST_Series_Calibration_Software.Zip was unzipped.
4.
In the lower left of the window, select the appropriate COM and
press
Open
.
For example:
5.
When connected, the numerical count to the right of
Sent:
will
increase in value:
.
This is the indication that the communication has been
established with the treadmill.