HEAT DISSIPATION DETAILS

DESCRIPTION

ROOM

MAX W

MAX BTU

Magnet (MAG) and Patient Table (PT)

Magnet Room

2400

8189

Blower Box (MG6)

Magnet Room

1000

3415

Magnet Monitor (MON)

Control/Equipment Room

60

205

System Cabinet (SC)

Control/Equipment Room

5000

17000

Operator Workspace with LCD Color Display (GOC)

Control/Equipment Room

1450

4947

Water Chiller for BRM (4kW LCS) (WC1)

Control/Equipment Room

1670

5695

Water Chiller for SC (8kW LCS)  (WC2)

Control/Equipment Room

1670

5695

Shield/Cryo Cooler Compressor - Water Cooled  (CRY)

Control/Equipment Room

500

1706

Options

GE pre-engineered Main Disconnect Panel (MDP)

Control/Equipment Room

264

900

Magnetic Resonance Elastography (MRE)

Equipment Room

200

682

TEMPERATURE AND HUMIDITY SPECIFICATIONS

IN-USE CONDITIONS

NOTE
In case of using air conditioning systems that have a risk of water leakage it is recommended not to install it above
electric equipment or to take measures to protect the equipment from dropping water.

AIR RENEWAL

According to local standards.

AREA

Temperature

Humidity

Maximum

Room Gradient

Range

Change /Hr

Range %

Change %

EQUIPMENT ROOM at Inlet

15-28 *

59-82.4 *

3

5

30-75 *

5

3**

5**

MAGNET ROOM

15-21

59-69.8

3

5

30-60 *

5

3

5

CONTROL ROOM

15-32

59-89.6

3

5

30-75 *

5

3

5

OUTDOOR (11kw Chiller)

-30-43

-22-110

-

-

5-100

-

-

-

NOTE

** Room temperature gradient specification applies from floor to height of top discharge of equipment cabinets.

The altitude is from 30.5 m [100 ft] below sea level to 2438 m [7992 ft] above sea level.

MAGNET ROOM EXHAUST FAN SCHEMATIC

Vent to outside

environment

Exhaust Fan

To Facility
Air Handler

AC Power (rated as required
for operation of Motorized

Damper and Exhaust Fan)
Ground secondary of low

voltage transformer to RF

Room Common Ground Stud

RF Shield

Dielectric Isolator

Room ceiling

Motorized Damper

Exhaust Intake Vent

MAGNET ROOM

CONTROL ROOM

RF Filter

ON/OFF

Manual

fan switches in

parallel

Ductwork

Ductwork

MAGNET ROOM VENTING REQUIREMENTS

HVAC VENT REQUIREMENTS

HVAC vendor must comply with Magnet room temperature and humidity specifications and RF shielding specifications.
RF Shield vendor must install open pipe or honeycomb HVAC waveguides.

All serviceable parts in the Magnet room (e.g.: diffusers) must be non-magnetic.
Waveguides must be nonmagnetic and electrically isolated.

Incoming air must contain at least 5% air from outside the Magnet room (inside or outside the facility) to displace
residual helium.

EMERGENCY VENT REQUIREMENT

Exhaust vent system is supplied by the customer.

All items within the RF enclosure must be non-magnetic.
The exhaust vent system must be tested and operational before the magnet is installed.
The exhaust intake vent must be located near the magnet cryogenic vent at the highest point on the finished or drop
ceiling.
The Magnet room exhaust fan and exhaust intake vent must have a capacity of at least 1200 CFM (

) with a

minimum of 12 room air exchanges per hour.

The exhaust fan must be placed above RF shielding located outside 10 gauss (1mT) and with appropriate waveguide.
The system must have a manual exhaust fan switch near the Operator Workspace and in the Magnet room near the
door (the switches must be connected in parallel).
All system components must be accessible for customer inspection, cleaning and maintenance

PRESSURE VENT REQUIREMENT

pressure equalizing vent is required in the magnet room ceiling or in the wall, at the highest point possible.
The vent minimum size must be (610 mm x 610 mm [24 in x 24 in]) or equivalent.
The pressure equalization vent must be located so any Helium gas is not vented into occupied areas.

Note: Location may affect acoustic noise transmission into occupied spaces.

Date

Rev

/24

EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG

A

03/Aug/2018

SIGNA EXPLORER /SIGNA CREATOR

Typical

M2 - HVAC-Venting

15

Содержание SIGNA EXPLORER

Страница 1: ...mplete set of final issue drawing GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings however caused All dimensions are in millimeters unless otherwise spe...

Страница 2: ...licable Power for MR compressor Chiller is available Ensure cryogen venting system is available for magnet connection Exhaust fan system is installed and operational per GE requirements Status of work...

Страница 3: ...ned to the magnet room If not possible a barrier is recommended to prevent entry to the 5g field area 2 The spaces around above and below the magnet must be reviewed for effects of the 5g 3g 1g and 5g...

Страница 4: ...contingent on a 72 in corridor width C 26 Minimum opening for is 43 in w x 82 in h contingent on a 96 in corridor width C 27 Counter top for equipment provide grommeted openings as required to route c...

Страница 5: ...efine RF shield s inset according to provisions made by the RF Shield vendor Warning 5 Gauss line outside the Magnet room limits MAGNET ROOM CONTROL ROOM FRONT VIEW WITH MAGNETIC FIELD 6 4 3 2 1 1 3 1...

Страница 6: ...ency generators Commercial laundry equipment Food preparation area Water cooling equipment HVAC equipment Major mechanical equipment room Credit cards watches and clocks Air conditioning equipment Fue...

Страница 7: ...net cover is adjustable between 100 200mm 4 8 in D PP cover is adjustable between 300 400mm 11 8 15 7 in SYSTEM CABINET Top View PENETRATION PANEL Top View D D PENETRATION PANEL Front View Cut out cen...

Страница 8: ...0 4 in 401 15 79 in 750 29 52 in 700 27 56 in 377 14 84 in 330 12 98 in 208 8 19 in GLOBAL OPERATOR CABINET GOC Center of Gravity SCALE 1 10 TOP VIEW FRONT VIEW SIDE VIEW 300 11 8 in PENETRATION PANEL...

Страница 9: ...ravity GAS SUPPLY RETURN WATER OUT IN 232 9 13 in 450 17 72 in 103 4 06 in 276 10 87 in 153 6 02 in 450 17 72 in 800 31 5 in 153 6 02 in 153 6 02 in BLOWER BOX A Air in take clearance B Duct trim cove...

Страница 10: ...ROUTING Floor must be able to withstand a moving load of 5320 daN Minimum Opening Height 2 5m 98 5 in width 2 5m 98 5 in Maximum slope 30 degree STRAIGHT PATH Rigging wheels required PATH WITH 90 DEGR...

Страница 11: ...safety all removable items within the magnet room such as faucet handles drain covers switch box cover plates light fixture components mounting screws etc Must be non magnetic If you have a specific...

Страница 12: ...ng 4 Structural wall backing for Magnet Rundown Unit 5 Structural wall backing for Main Disconnect Panel 6 Structural wall backing for Magnet Monitor 7 Floor levelness area for Systems Cabinet Date Re...

Страница 13: ...ded to the structural elements and not to common screed for magnet anchoring table dock anchor hole VibroAcoustic Pad weight 8 kg 17 lbs each Seismic anchor holes 1346 53 in 673 26 5 in 250 9 8 in 175...

Страница 14: ...supplied and installed by the customer or his contractors For complete design and requirements specifications and guidelines refer to the pre installation manual system cooling cryogen venting wavegu...

Страница 15: ...xhaust Intake Vent MAGNET ROOM CONTROL ROOM RF Filter ON OFF Manual fan switches in parallel Ductwork Ductwork MAGNET ROOM VENTING REQUIREMENTS HVAC VENT REQUIREMENTS HVAC vendor must comply with Magn...

Страница 16: ...ow 30 l min 8 gpm Temperature rise at maximum flow Temperature rise at minimum flow Maximum inlet pressure of Chiller Cryo Compressor 690 kPa 100 psi Minimum inlet pressure of Chiller Cryo Compressor...

Страница 17: ...PIPE DETAIL MAGNET VENT ADAPTER Do not remove or modify the vent adaptor bolted to the magnet VENT PIPE 610 mm 24 in may be cut to a minimum of 100 mm 4 in VENTGLAS CONNECTOR WAVEGUIDE Additional pipe...

Страница 18: ...10 63 4 52 31 17 8 22 56 68 2 26 15 58 4 11 28 34 10 in 250mm 0 20 0 6 1 0 03 0 68 0 55 3 79 0 82 5 65 0 27 1 86 0 04 2 83 20 40 6 1 12 2 0 07 1 58 0 82 5 65 1 51 10 41 0 41 2 83 0 75 5 17 40 60 12 2...

Страница 19: ...n direct as possible otherwise may result in the need for greater than standard cable lengths refer to the interconnection diagram for maximum usable lengths point to point 8 Conduit turns to have lar...

Страница 20: ...ephone lines network connection Duplex hospital grade dedicated outlet 120 v emergency single phase power 15a Duplex hospital grade dedicated outlet 120 v single phase outlet routed through RF filter...

Страница 21: ...P SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical E3 Electrical Elevations 21 MAGNET ROOM EQUIPMENT ROOM 8 9 TBD TBD 5 4 7 9 9 8 CONTROL ROOM 5 4 3 2 TBD TBD TBD TBD TBD TBD...

Страница 22: ...THE FLOOR NOT TO SCALE Removable coverplate DUCT ON THE WALL Date Rev 24 EN MRI TYP SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical E4 Electrical Details 22 CABLE MANAGEMENT...

Страница 23: ...E EMERGENCY OFF BUTTONS ARE SUPPLIED WITH GE MDP OPTION EMERGENCY OFF BUTTONS ARE CUSTOMER SUPPLIED IF GE MDP OPTION NOT USED CIRCUIT BREAKERS ARE PROVIDED FOR PDU CRYO COOLER COMPRESSOR CHILLER GRADI...

Страница 24: ...5 75m 13 94 51 67 ft 17 98 m 59 ft 9 56 m 31 ft 12 5 m 41 ft 12 8m 42 ft 17 4m 57 ft 2 55 16 13m 8 35 52 9 ft 12 2m 40 ft 21 00 m 68 ft 4 3 11 3m 14 1 37 1 ft GOC MAGNET ROOM CONTROL ROOM EQUIPMENT RO...