SYSTEMS CABINET & BLOWER BOX FLOOR REQUIREMENTS
SYSTEMS CABINET
FLOOR LEVELNESS
BLOWER BOX
ANCHORING
RF Screen
650
[25.6 in]
650
[25.6 in]
Specification of Floor
1.
Floor slope: < +/- 0.5 deg
2.
Floor surface: < +/-5mm
3.
Non-compressible flooring material
only, for example, no carpet allowed.
NOTE: The blower box should be anchored with
M6 bolts which can stand 0.217 Kilo Newton
shear force and 0.076 Kilo Newton tension force.
147.65
[5.81 in]
147.65
[5.81 in]
49.3
[1.94 in]
MAGNET ROOM FLOOR SPECIFICATIONS
Magnet, Enclosure, and Patient Table areas must be flat and level within 3 mm (0.125 in) within the shaded area
shown.
SCALE 1:50
The finished floor must support the weight of all components (e.g., patient table, gradient coil replacement cart)
throughout operation and service life.
1825
[71.9 in]
5532
[217.8 in]
NOT TO SCALE
MAGNET ON VIBROACOUSTIC DAMPENING KIT
Anchoring to the floor is
intended to the structural
elements and not to common
screed.
for magnet anchoring
table dock
anchor hole
VibroAcoustic Pad weight: 8 kg (17 lbs) (each)
Seismic anchor holes
1346
[53 in]
673
[26.5 in]
250
[9.8 in]
175
[6.9 in]
1354
[53.3 in]
76
[3 in]
PATIENT TABLE DOCK ANCHOR MOUNTING REQUIREMENTS
The RF Shield vendor must design and install the dock anchor bolt
The dock anchor hole must be drilled after the Magnet is installed
The dock anchor must not contact floor rebar or other structural steel
The dock anchor must electrically contact the RF shield at point of entry
The RF shield vendor must perform a pull test on the anchor (equal to the clamping force).
NOT TO SCALE
FEMALE ANCHOR METHOD
(CUTAWAY SIDE VIEW)
BOLTED METHOD
(CUTAWAY SIDE VIEW)
1
Removable Anchor Rod (Male insert)
2
Dock
3
Clamp bracket
4
Finished floor
5
Filler Board or Grout
6
RF Shield
7
Conductive Fibrous Washer (RF seal)
8
Concrete
9
Female Anchor Insert
1
2
3
4
5
6
7
8
9
8
7
6
5
4
3
2
1
Date
Rev
/24
EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG
A
03/Aug/2018
SIGNA EXPLORER /SIGNA CREATOR
Typical
S3 - Structural Details
13