TYPICAL CRYOGEN SIDE WALL EXIT WITH LONG SWEEP ELBOW
Mesh bird screen
38 (1.5 in)
insulation
rigid support
203 [8 in]
ventglas outside
RF room
KEY COMPONENTS :
RF waveguide extended from wall to magnet adapter.
Must be all same material and all welded.
Support system must withstand 8229 N (1850 lbs)
GE ventglass must be installed in vertical section directly over magnet
RF WAVEGUIDE EXTENDED
OUTSIDE
25
[1 in
]
TYPICAL CRYOGENIC VENT PIPE DETAIL
MAGNET VENT ADAPTER
Do not remove or modify the
vent adaptor bolted to the
magnet.
VENT PIPE
610 mm [24 in], may be cut to a
minimum of 100 mm [4 in].
VENTGLAS
CONNECTOR
WAVEGUIDE
Additional pipe/tubing
[0.5 in] mesh
DIELECTRIC
BREAK
CRYOGENIC VENTING ( EXTERIOR )
GROUND
VENT CAP
(customer supplied)
OUTSIDE WALL
ROOF
VENT CAP
(customer
supplied)
NOT TO SCALE
EXHAUST AREA
6.1m x 4.6m
[20ft x 15ft] (LxW)
EXHAUST AREA
6.1m x 4.6m
[20ft x 15ft] (LxW)
4600mm
[181.1in]
Exclusion area
Warning sign
914.4mm
[36in]
3657.6mm
[144in]
Date
Rev
/24
EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG
A
03/Aug/2018
SIGNA EXPLORER /SIGNA CREATOR
Typical
M4 - Cryogenics (1)
17