POWER REQUIREMENTS

INTRODUCTION

The system requires two independent power inputs:

main power supply
Magnet monitor power

SPECIFICATIONS OF MAIN POWER INPUT

Power input must be separated from any others which may generate transients (elevators, air conditioning,
radiology rooms equipped with high speed film changers...).

Total harmonic distortion less than 2.5%.

Phase imbalance must not exceed 2%.

SPECIFICATIONS OF MAGNET MONITOR POWER

Power Supply (3
G)

Frequency 50Hz/60Hz

480

415

400

380

208

200

Total Current

60.1

66.8

69.3

72.9

N/A

N/A

Maximum input
power (5 sec max)

59.15 kVA

Stand-by power

13.4 kVA at 0.9 lagging Power Factor including 4.4 KVA for PDU

and 9KVA (continuous operation) for Shield/Cryo Cooler Cabinet.

Critical Power Requirements is different per each configuration.

POWER INPUT

EMERGENCY LIFE-SAFETY POWER SINGLE PHASE + N +G

POWER DEMAND

2.0 A

VOLTAGE

110/220V

FREQUENCY

50 / 60 Hz

CABLES

Power and cable installation must comply with the distribution diagram.
Size of the Main power input cable is determined by the customer, taking its length and admissible voltage
drops  into consideration.

All cables must be isolated and flexible, cable color codes must comply with standards for electrical
installation.
The cables from signaling and remote control (Y,Emergency Off Buttons,L...) will go to the Distribution

Panel/Box with a pigtail length of 1.5m [60in], and will be connected during installation.
Each conductor will be identified and isolated (screw connector).

GROUND SYSTEM

The equipotential link will be by means of an equipotential bar.
The grounding point of Distribution Panel/Box is directly connected to the building's ground by an isolated
copper cable.
The impedance of the earth bar should be less than or equal to 2 ohms

MAIN DISCONNECT PANEL (MDP)

NEUTRAL

GROUND

NEUTRAL MUST

BE TERMINATED
INSIDE THE MDP

NEUTRAL
BUSS BAR

G/PE
BUSS

BAR

CB#1

3-Black
1-White
1-Green

FACILITY

INPUT

POWER

(480-V,

3 Phase)

Cable SUPPLIED BY CUSTOMER

Equipment SUPPLIED BY GE

Equipment SUPPLIED BY
CUSTOMER

Cable SUPPLIED BY GE

POWER DISTRIBUTION

NOTES:

RUNS 296 AND 297, & POWER CORDS FOR SHIELD/CRYO COOLER
COMPRESSOR CABINET, MRCC, & MAGNET MONITOR EQUIPMENT (MAGNET
MONITOR, UPS INPUT & OUTPUT, MODEM, OPTIONAL MULTIPLEXER) ARE
GE SUPPLIED CABLES. ALL OTHER WIRING IS CUSTOMER SUPPLIED.

TWO REMOTE EMERGENCY "OFF" BUTTONS ARE SUPPLIED WITH GE MDP
OPTION, EMERGENCY OFF BUTTONS ARE CUSTOMER SUPPLIED IF GE MDP
OPTION NOT USED.

CIRCUIT BREAKERS ARE PROVIDED FOR PDU, CRYO COOLER COMPRESSOR
CHILLER, GRADIENT CHILLER, SHIELD/CRYO COOLER COMPRESSOR CABINET,
MAGNET MONITOR EQUIPMENT CIRCUITS.

ALL BRANCH CIRCUITS DROP OUT ON LOSS OF POWER. COLDHEAD MRCC,
GRADIENT MRCC, SHIELD/CRYO COOLER COMPRESSOR CABINET, & MAGNET

MONITOR EQUIPMENT AUTOMATICALLY RESTART AFTER 3 SEC TIME DELAY
UPON RESTORATION OF POWER. EMERGENCY OFF LOCKS OUT ALL
CONTACTORS.

IF 3 PHASE WYE WITH NEUTRAL AND GROUND (5 WIRE SYSTEM) INPUT USED

THEN NEUTRAL MUST BE TERMINATED INSIDE THE MAIN DISCONNECT

PANEL AND NOT BROUGHT TO THE POWER CABINET.

SUPERVISORY CIRCUIT FOR HVAC INTERLOCK CONTACTS OPEN ON LOSS OF
DC POWER OR EMERGENCY OFF POSITION.

EQUIPMENT ROOM

"EMERGENCY OFF"

BUTTON

MAGNET ROOM

"EMERGENCY
OFF" BUTTON

PENETRATION

PANEL (PP1)

ARM = AUTO

RESTART MODULE

SHIELD/CRYO COOLER

COMPRESSOR CABINET

(MS5)

GND

FACILITY POWER (24
HOURS/DAY, 7 DAYS/WEEK)
RECOMMENDED TO MAXIMIZE
PROPER UNINTERRUPTED
MAGNET OPERATION FOR
SHIELD/CRYO COOLER
COMPRESSOR CABINET & CRYO
COMPRESSOR CHILLER (MRCC) IF
SUPPLING WATER COOLING FOR
SHIELD/CRYO COOLER
COMPRESSOR CABINET

GROUND STUD OF
SYSTEM CABINET

TO TERMINAL STRIP IN PDU

MODULE LOCATED

BOTTOM OF SYSTEMS

CABINET

POWER DISTRIBUTION UNIT

(PD1)

SYSTEM CABINET

SPLICE

ARM

ARM

CB #3

CB #5

CB #2

EMER-GENCY
OFF MODULE

SUPERVISORY

CKT FOR HVAC

INTERLOCK
CONTACTS

AUTO RESTART

CONTROL

CIRCUIT

120
VAC

FUSES

ARM

CB #3

A

B

C

{

NC

NC

RF SHIELD

RUN 296

RUN 297

BLK

BLK

RED

RED

CB AMPS

1

150

2

100

3

20

4

20

5

30

6

15

7

35

Date

Rev

/24

EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG

A

03/Aug/2018

SIGNA EXPLORER /SIGNA CREATOR

Typical

E5 - Power Requirements

23

FEEDER TABLE

MIN. feeder wire size,

AWG or MCM (sq. M)/VAC

Minimum feeder wire length - ft (m)

100 (30.5)

150 (46)

200 (61)

250 (76)

300 (92)

350 (107)

400 (122)

450 (137)

480 VAC

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

1/0 (70)

Ground req'd

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

6 (17)

GENERAL NOTES

In all cases qualified personnel must verify that the feeder (at the point of take-off) and the run to the MR system meet all the requirements

stated in the PIM

For a single unit installation, the minimum transformer size is 112.5KVa.  Regulated transformer is not required unless voltage changes

/- 10% over a period of 1 hour or longer

Grounding conductor will run from the equipment back to  the power source/main grounding point and always travel in the same conduit

with the feeders

Содержание SIGNA EXPLORER

Страница 1: ...mplete set of final issue drawing GE cannot accept responsibility for any damage due to the partial use of GE final issue drawings however caused All dimensions are in millimeters unless otherwise spe...

Страница 2: ...licable Power for MR compressor Chiller is available Ensure cryogen venting system is available for magnet connection Exhaust fan system is installed and operational per GE requirements Status of work...

Страница 3: ...ned to the magnet room If not possible a barrier is recommended to prevent entry to the 5g field area 2 The spaces around above and below the magnet must be reviewed for effects of the 5g 3g 1g and 5g...

Страница 4: ...contingent on a 72 in corridor width C 26 Minimum opening for is 43 in w x 82 in h contingent on a 96 in corridor width C 27 Counter top for equipment provide grommeted openings as required to route c...

Страница 5: ...efine RF shield s inset according to provisions made by the RF Shield vendor Warning 5 Gauss line outside the Magnet room limits MAGNET ROOM CONTROL ROOM FRONT VIEW WITH MAGNETIC FIELD 6 4 3 2 1 1 3 1...

Страница 6: ...ency generators Commercial laundry equipment Food preparation area Water cooling equipment HVAC equipment Major mechanical equipment room Credit cards watches and clocks Air conditioning equipment Fue...

Страница 7: ...net cover is adjustable between 100 200mm 4 8 in D PP cover is adjustable between 300 400mm 11 8 15 7 in SYSTEM CABINET Top View PENETRATION PANEL Top View D D PENETRATION PANEL Front View Cut out cen...

Страница 8: ...0 4 in 401 15 79 in 750 29 52 in 700 27 56 in 377 14 84 in 330 12 98 in 208 8 19 in GLOBAL OPERATOR CABINET GOC Center of Gravity SCALE 1 10 TOP VIEW FRONT VIEW SIDE VIEW 300 11 8 in PENETRATION PANEL...

Страница 9: ...ravity GAS SUPPLY RETURN WATER OUT IN 232 9 13 in 450 17 72 in 103 4 06 in 276 10 87 in 153 6 02 in 450 17 72 in 800 31 5 in 153 6 02 in 153 6 02 in BLOWER BOX A Air in take clearance B Duct trim cove...

Страница 10: ...ROUTING Floor must be able to withstand a moving load of 5320 daN Minimum Opening Height 2 5m 98 5 in width 2 5m 98 5 in Maximum slope 30 degree STRAIGHT PATH Rigging wheels required PATH WITH 90 DEGR...

Страница 11: ...safety all removable items within the magnet room such as faucet handles drain covers switch box cover plates light fixture components mounting screws etc Must be non magnetic If you have a specific...

Страница 12: ...ng 4 Structural wall backing for Magnet Rundown Unit 5 Structural wall backing for Main Disconnect Panel 6 Structural wall backing for Magnet Monitor 7 Floor levelness area for Systems Cabinet Date Re...

Страница 13: ...ded to the structural elements and not to common screed for magnet anchoring table dock anchor hole VibroAcoustic Pad weight 8 kg 17 lbs each Seismic anchor holes 1346 53 in 673 26 5 in 250 9 8 in 175...

Страница 14: ...supplied and installed by the customer or his contractors For complete design and requirements specifications and guidelines refer to the pre installation manual system cooling cryogen venting wavegu...

Страница 15: ...xhaust Intake Vent MAGNET ROOM CONTROL ROOM RF Filter ON OFF Manual fan switches in parallel Ductwork Ductwork MAGNET ROOM VENTING REQUIREMENTS HVAC VENT REQUIREMENTS HVAC vendor must comply with Magn...

Страница 16: ...ow 30 l min 8 gpm Temperature rise at maximum flow Temperature rise at minimum flow Maximum inlet pressure of Chiller Cryo Compressor 690 kPa 100 psi Minimum inlet pressure of Chiller Cryo Compressor...

Страница 17: ...PIPE DETAIL MAGNET VENT ADAPTER Do not remove or modify the vent adaptor bolted to the magnet VENT PIPE 610 mm 24 in may be cut to a minimum of 100 mm 4 in VENTGLAS CONNECTOR WAVEGUIDE Additional pipe...

Страница 18: ...10 63 4 52 31 17 8 22 56 68 2 26 15 58 4 11 28 34 10 in 250mm 0 20 0 6 1 0 03 0 68 0 55 3 79 0 82 5 65 0 27 1 86 0 04 2 83 20 40 6 1 12 2 0 07 1 58 0 82 5 65 1 51 10 41 0 41 2 83 0 75 5 17 40 60 12 2...

Страница 19: ...n direct as possible otherwise may result in the need for greater than standard cable lengths refer to the interconnection diagram for maximum usable lengths point to point 8 Conduit turns to have lar...

Страница 20: ...ephone lines network connection Duplex hospital grade dedicated outlet 120 v emergency single phase power 15a Duplex hospital grade dedicated outlet 120 v single phase outlet routed through RF filter...

Страница 21: ...P SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical E3 Electrical Elevations 21 MAGNET ROOM EQUIPMENT ROOM 8 9 TBD TBD 5 4 7 9 9 8 CONTROL ROOM 5 4 3 2 TBD TBD TBD TBD TBD TBD...

Страница 22: ...THE FLOOR NOT TO SCALE Removable coverplate DUCT ON THE WALL Date Rev 24 EN MRI TYP SIGNA_EXPLORER WEB DWG A 03 Aug 2018 SIGNA EXPLORER SIGNA CREATOR Typical E4 Electrical Details 22 CABLE MANAGEMENT...

Страница 23: ...E EMERGENCY OFF BUTTONS ARE SUPPLIED WITH GE MDP OPTION EMERGENCY OFF BUTTONS ARE CUSTOMER SUPPLIED IF GE MDP OPTION NOT USED CIRCUIT BREAKERS ARE PROVIDED FOR PDU CRYO COOLER COMPRESSOR CHILLER GRADI...

Страница 24: ...5 75m 13 94 51 67 ft 17 98 m 59 ft 9 56 m 31 ft 12 5 m 41 ft 12 8m 42 ft 17 4m 57 ft 2 55 16 13m 8 35 52 9 ft 12 2m 40 ft 21 00 m 68 ft 4 3 11 3m 14 1 37 1 ft GOC MAGNET ROOM CONTROL ROOM EQUIPMENT RO...

Отзывы: