4. Conventional Operation: Menu System, Keypad, and Displays
PHASOR XS Operating Manual
Page 57
any time a probe of a different type is installed.
Selecting Probe Type
(SETUP-RECEIVER-DUAL)
Step 1:
Activate the RECEIVER Submenu (located in the
SETUP Menu by pressing
below it.
Step 2:
Press
next to the selection titled DUAL.
Step 3:
To change the probe type, continue pressing
or turn the function knob. Each available probe type is rep-
resented by an icon that’s displayed in the Icon Bar (near
the upper left corner of the display) whenever that probe
type is indicated. The following types are available:
• ON—For dual-element probes (
will be
displayed)
• OFF—For single-element probes (
will be
displayed)
Step 4:
The probe type will be set to the last one
displayed.
Specifying the Probe Frequency
(SETUP-RECEIVER-FREQUENCY)
Step 1:
Activate the RECEIVER Submenu (located in the
SETUP Menu) by pressing
below it.
Step 2:
Press
next to the selection titled
FREQUENCY.
Step 3:
To change the specified frequency, continue
pressing
or turn the function knob. You’ll note that the
following frequency settings are available:
•
1, 2, 2.25, 4, 5, 10, 13 MHz
—Set to match conven-
tional probe frequencies
•
BROADBAND
—Select to utilize a built-in broad-
band (BB) filter
Step 3:
The probe frequency level will be set to the last
one displayed.
Modifying the Signal Ratio to Noise by
Changing the Damping Level
(SETUP-PULSER-DAMPING)
Step 1:
Activate the PULSER Submenu (located in the
SETUP Menu) by pressing
below it.
Step 2:
Press
next to the selection titled DAMPING.
Step 3:
To change the specified damping level and op-
timize the A-Scan signal appearance, continue pressing
or turn the function knob. You’ll note that the following
damping levels are available:
•
50 or 1000
Ω
(see
Figure 4-9
)
Step 4:
The damping level will be set to the one last
displayed.
4.7.3 Adjusting the Pulser Repetition
Frequency
(PRF)
The Pulser fires at a frequency which can either be set
automatically or manually. To set the PRF mode and
frequency level
Step 1:
Activate the SETUP Menu (located in the HOME
Menu) by pressing
below it.
Step 2:
Select the PULSER Submenu by pressing
FIGURE 4-8—Probe Attachment
Locations
Содержание Phasor 16/16
Страница 6: ...Contents Page 4 PHASOR XS Operating Manual ...
Страница 16: ...1 Phased Array Operation Keypad Menus and Displays Page 14 PHASOR XS Operating Manual ...
Страница 24: ...2 Setting Up for Phased Array Measurement Page 22 PHASOR XS Operating Manual ...
Страница 45: ...3 Operating in Phased Array Mode PHASOR XS Operating Manual Page 43 FIGURE 3 9 TCG Recording Process ...
Страница 86: ...5 Making Conventional Measurements Page 84 PHASOR XS Operating Manual ...
Страница 92: ...6 Storing Data Sets and Generating Reports Page 90 PHASOR XS Operating Manual ...
Страница 96: ...7 Technical Specifications Page 94 PHASOR XS Operating Manual ...