5. Making Conventional Measurements
PHASOR XS Operating Manual
Page 83
Step 2:
Adjust the TRANSFER CORR function as required
to compensate for coupling differences.
NOTE:
While TRANSFER CORR is set to any value
other than 0, the displayed gain value will appear in a
contrasting color, indicating that the amount of applied
gain differs from the listed “instrument” gain (the con-
trasting color does not appear when the display color
is set to the high-contrast BLACK on WHITE).
5.17.3 Deleting a JISDAC Curve
To delete a stored JISDAC curve
Step 1:
With the JISDAC Menu activated, select the
SETUP Submenu.
Step 2:
Press
next to the DELETE CURVE function
(see
Figure 5-6
).
Step 3:
Press
next to the DELETE CURVE function a
second time. Then press
to confirm your selection.
Step 4:
The statement in the JISDAC function box will
change to OFF.
FIGURE 5-14—Storing at least two JISDAC points allows you to generate a recorded reference line (identified as H) and
corresponding offset lines. Line H, M, or L can be chosen as the reference line from which measurements are based.
Содержание Phasor 16/16
Страница 6: ...Contents Page 4 PHASOR XS Operating Manual ...
Страница 16: ...1 Phased Array Operation Keypad Menus and Displays Page 14 PHASOR XS Operating Manual ...
Страница 24: ...2 Setting Up for Phased Array Measurement Page 22 PHASOR XS Operating Manual ...
Страница 45: ...3 Operating in Phased Array Mode PHASOR XS Operating Manual Page 43 FIGURE 3 9 TCG Recording Process ...
Страница 86: ...5 Making Conventional Measurements Page 84 PHASOR XS Operating Manual ...
Страница 92: ...6 Storing Data Sets and Generating Reports Page 90 PHASOR XS Operating Manual ...
Страница 96: ...7 Technical Specifications Page 94 PHASOR XS Operating Manual ...