Section 5.0 - Gantry Cable Connections
5.2
Cardiac Monitor Setup
Note:
The following procedure is applied only if the Accessory I/F Hardware (IPC) Option is installed.
1.) Follow install instructions shipped with monitor to set up the monitor stand with basket.
2.) After the stand is assembled, mount the monitor to the stand by sliding the monitor onto the
plate.
3.) Pull down the front pin on plate and slide monitor until it snaps into place.
4.) Secure the monitor using the two nylon set screws under the plate.
5.) Connect the long CAT5 LAN cable between back side of Gantry option interface and Console
EKG port.
The long CAT5 cable is included in IVY Monitor kit.
6.) Attach the cables.
Do not use the cables shipped with the monitor
, find the 5317480 cable
included with the Cardiac Cable Kit (E8007TB).
CAUTION
The outlets are not for General Use. Gantry outlets have a rating for 3.0A at 120VAC.
Accessories should not exceed above rating.
7.) Connect the IEC power cord, ground wire, LEMO and CAT5 to the gantry option interface
panel. (See
Figure 2-21 Gantry Option Interface Panel
Figure 2-22 Connections on Rear of Cardiac Monitor (IVY3150)
Cardiac Monitor Power
Ground
Cat 5
LEMO
Cardiac Monitor Power
Ground
(Alternative)
(Alternative)
Left Panel
Right Panel
If the wire power cord cable
clamp is too short, replace it
with the supplied longer clamp.