Chapter 2 - Power, Ground & Interconnect Cables
Page 113
2 –
Install
Power
3.6
AC Box Connections
CAUTION
The outlets are not for General Use. Operator Console outlet has a rating for 2.5A at 120VAC.
Accessories should not exceed above rating.
Note:
Console power is single phase power. Outlet assigned is not critical.
1.) Connect the console power cable and ground cable to the console power panel.
2.) Connect console component power cords as listed in
. (“J numbers” increment from
top to bottom, left to right)
Figure 2-16 AC Box Connections
Number
Description
J9
Display Monitor Power Connection
J10
Scan Monitor Power Connection
J11
Peripheral Media Tower Power Connection
J12
In-Room Monitor Connection
J13
Injector Power Connection
J14
RPM Power Connection
J56
GSCB Power Connection
Table 2-10 AC Box Outlet Assignments
CB3
J13
J14
J55
JI-J6 & J9-J12 controlled by CB1;
J13, J14, J55 controlled by CB3.
J57
120V AC 2.5A 50/60HZ
J56
J10
J11
J12
J9
120V AC 10A 50/60HZ
12V DC
120V AC 10A 50/60HZ
120V AC 10A 50/60HZ
120V AC 10A 50/60HZ
120V AC 2.5A 50/60HZ
120V AC 2.5A 50/60HZ