5
Trouble-shooting
30
Monitor UV-900 User Manual 18-1120-05 Edition AF
5.3
Error messages
If the suggested actions do not correct the fault, call GE Healthcare.
Message
Action
Sample fiber failure
There may be too high absorption in the UV cell.
1. Check the liquid.
2. Ensure that there are no air bubbles in the system.
3. Check the connections of the UV cell optical fibers.
4. Clean the UV cell, see section
Cleaning the flow cell and
optical fiber connectors
.
Not calibrated
Calibration failed
Recalibrate by switching the instrument off and then back to
on, or choose
Recalibrate
in the
Check
menu
Change lamp
Call GE Healthcare for Xe-lamp exchange
Mode fiber failure
Reference fiber failure
Lamp module failure
Low light intensity
Block filter is defect
Call GE Healthcare
High intensity
in Reference fiber
Recalibrate by switching the instrument off and on
High intensity
in Sample fiber
1. Disconnect the optical fiber connectors.
2. Recalibrate by switching the instrument off and on.
ERROR Number 10
ERROR Number 16
ERROR key(OK)
ERROR key(Esc)
ERROR key(OK+Esc)
ERROR 100
ERROR 109-113
ERROR 120-121
1. Switch off the instrument.
2. Check all connections.
3. Switch on the instrument.
ERROR 106-108
ERROR 118
1. Switch off the instrument.
2. Check all UniNet connections.
3. Switch on the instrument.