Reference information
C
Monitor UV-900 User Manual 18-1120-05 Edition AF
39
C
Technical specifications
Operating data
Wavelength range
190–700 nm in steps of 1 nm,
3 wavelengths simultaneously
Bandwidth
4 nm
Wavelength accuracy
±2 nm
Wavelength reproducibility
±0.01 nm
Wavelength switch time
< 500 ms (one cycle from 214 nm to
254 nm and back to 214 nm)
Linearity
< 2% deviation up to 2 AU at 260 nm with
Uracil at pH 2
Noise (specified for 10 mm
flow cell)
1, 2
Single wavelength
short term (0.5–1 min)
long term (1–10 min)
Dual wavelengths
short term (0.5–1 min)
long term (1–10 min)
< 6x10
-5
AU at 230 nm
< 6x10
-5
AU at 230 nm
< 2x10
-4
AU at 230 and 254 nm
< 2x10
-4
AU at 230 and 254 nm
Drift
2)
< 2x10
-4
AU/h at 254 nm
Environment
+4 to +40 °C
20–95% relative humidity
84–106 kPa (840-1060 mbar)
atmospheric pressure
1)
Measured with water at 1 ml/min, time constant 1 s.
2)
Typical values at room temperature after varm-up.
Flow cell
Max. flow rate
100 ml/min
Max. pressure
2 MPa (20 bar, 290 psi)
Backpressure
Max. 0.5 bar at 2 ml/min with water at
25 °C
Liquid temperature range
+4 to +40 °C
Optical path length
2 mm cell
10 mm cell
2 mm
10 mm
Cell volume
2 mm cell
10 mm cell
2 µl
8 µl
Degree of protection
IP 21
Wetted materials
PTFE (polytetrafluoroethylene)
PEEK (polyetheretherketone)
Titanium (palladium alloy)
Quartz (synthetic fused silica)
pH stability range
1–13, 1–14 (<1 days exposure)