Operation
3
Monitor UV-900 User Manual 18-1120-05 Edition AF
19
3.7
Autozero
The autozero function sets the detected absorbance to zero when the
OK
button is pressed. All three wavelengths are autozeroed. Autozero is
recommended after wavelength changes in a method and before the sample
is injected.
1 Select main menu
Autozero
, press
OK
. The normal absorbance value
display is then shown.
In UNICORN, Autozero is set with the instruction
AutoZeroUV
in
System
Control:Manual:Alarm&Mon
.
3.8
Storage and shut-down
Overnight:
The flow cell can be left filled with buffer.
Weekend and long term storage:
Flush the flow cell with distilled water and
then fill it with 20% ethanol.
The flow cell can also be stored dry by flushing as above with distilled water
and then blowing a compressed inert gas such as nitrogen (N
2
) through the cell.
Replace the red protective caps. Never use compressed air as this may contain
droplets of oil.
Autozero
CAUTION!
Do not allow solutions which contain dissolved salts, proteins or
other solid solutions to dry out in the flow cell. Do not allow particles to
enter the flow cell as damage to the flow cell may occur.