Installation
12
Monitor UV-900 User Manual 18-1120-05 Edition AF
2.4
Connecting electrical signal cables
The sockets for electrical signals are located on the back panel.
2.4.1
Connecting to recorder (if used)
1 Connect the recorder to the miniDIN-socket
Analogue out
using the cable
supplied. Each wavelength signal is available on a separate channel using
the following wires:
Note:
The signal cable is delivered with protective covers on each wire. Do
not remove the protective covers from unused connections as a
short circuit may disturb the measurements.
2 Set the recorder to 0–1 V input, full scale, 0 V offset.
2.5
Connecting to communication link
The monitor can be used in ÄKTAexplorer or ÄKTApurifier and controlled from a
PC running UNICORN
™
version 3.21 or higher, using UniNet cables.
Analogue out
0-1V
Remote UniNet 1
Mains
On/off
λ
1
Wire 1
+
Wire 2
–
λ
2
Wire 3
+
Wire 4
-
λ
3
Wire 5
+
Wire 6
-
CAUTION!
The mains power to the ÄKTAexplorer or ÄKTApurifier must be
switched OFF before connecting the instrument to the UniNet-1 link.
WARNING!
Only use mains cables delivered or approved by GE Healthcare.
WARNING!
Do not block the rear panel of the system. The mains power
switch must always be easy to access.