3-4
MAC
TM
600
Revision G
2047426-002
Troubleshooting:
General Fault Isolation
Visual Inspection
A visual inspection of the equipment can detect disconnected cables,
debris on circuit boards, missing hardware, and loose components. These
could cause symptoms, and equipment failures.
NOTE
Perform the visual checks before starting any troubleshooting
procedures.
Table 3-1. Visual Inspection List
Sl No:
Area
Look for:
1
I/O connectors and
cables and AC power
cord
Fraying or other damage
Bent prongs or pins
Cracked housing
Loose screws in plugs
2
Interface cables
Excessive tension or wear
Loose connection
Strain reliefs out of place
3
Circuit boards
Moisture, dust, or debris (top and bottom)
Loose or missing components
Burn damage or smell of over-heated
components
Socketed components not firmly seated
PCB not seated properly in edge connectors
Solder problems: cracks, splashes on board,
incomplete feed through, prior modifications or
repairs
4
Ground wires/wiring
Loose wires or ground strap connections
Faulty wiring
Wires pinched or in vulnerable position
5
Fasteners
Loose or missing screws or other hardware,
especially fasteners used as connections to ground
planes on PCBs
6
Power source
Faulty wiring, especially AC outlet
Circuit not dedicated to system
(Power source problems can cause static
discharge, resetting problems, and noise.)
7
Keypad
Cuts or cracks in keypad membrane
Illegible labels
8
LCD display
Scratched or opaque display filter
9
Battery pack
Cracked, swollen, or leaky battery pack
enclosure
Debris on battery pack electrical contacts
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...