Maintenance:
Preparing System for FRU Replacement
Earth Cable Routing
Verify the earth cable routing is as shown in the following figure. It
should connect screw A and screw B. This is the Class II Earth (ground)
cable routing. To change the cable routing to comply with Class I
requirements, perform the following procedure.
1. Open the device and printer door as described in
2. Remove the screws from location A and B.
3. Remove the earth cable (2033817-001).
4. Using the cable and screws from steps 2 and 3, place the cable on
location C and D. The other end D can be connected to either of the
screw points on the printer base. Ensure the cable does not touch any
components on the PCB board and secure it in place with the screws.
Table 4-7. Routing the Earth Cable
Legend
Description
1
Assembly of the earthing cable with main
PCB assembly
2
Earthing cable assembly location
A and B
Class II Earth cable routing
C and D
Class I Earth cable routing
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...