Revision G
MAC
TM
600
4-29
2047426-002
Maintenance:
Check out Procedure
Verify the current leakage test results meet the requirements. Perform
electrical safety checks when indicated. All indicated electrical safety
checks require a pass/ fail indication for the steps performed. Record the
measurement values in your debrief.
1. Verify the equipment does not propose any health hazard. See
Leakage (AC line) Current Test”
on page 4-26, for more information.
2. Verify all exposed metal is properly grounded. See
on page 4-24 for more information.
NC = Normal Condition; SFC = Single Fault Condition; N/A = Not
Applicable
UUT = Unit Under Test
Table 5. Electrical Safety Checks
Step
Condition
UUT - ON
Result
Leakage Current Limits
Earth Leakage Current
1
Forward Polarity
NC
_______uA
Pass/Fail
150 uA
2
Neutral Open, Forward Polarity
SFC
_______uA
Pass/Fail
1000 uA
3
Neutral Open, Reverse Polarity
SFC
_______uA
Pass/Fail
1000
uA
4
Reverse Polarity
NC
_______uA
Pass/Fail
150
uA
Patient Leakage Current To Ground
1
Forward Polarity
NC
_______uA
Pass/Fail
10
uA
2
Neutral Open, Forward Polarity
SFC
_______uA
Pass/Fail
50 uA
3
Ground Open, Forward Polarity
SFC
_______uA
Pass/Fail
50 uA
4
Ground Open, Reverse Polarity
SFC
_______uA
Pass/Fail
50 uA
5
Neutral Open, Reverse Polarity
SFC
_______uA
Pass/Fail
50 uA
6
Reverse Polarity
NC
_______uA
Pass/Fail
10
uA
Ground Resistance
1
AC mains power cord ground prong
to exposed metal surface (ground lug)
N/A
_______
Q
Pass/Fail
Less than 1 Ohm
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...