Revision G
MAC
TM
600
4-27
2047426-002
Maintenance:
Check out Procedure
The test has failed if the measured values are greater than:
Patient Leakage Current, Mains on Applied Part
WARNING
PERSONAL SAFETY — MAINS VOLTAGE is applied to
floating input (patient connection) and patient cables/test
plug conductors and safety tester PATIN JACK. For the
protection of the person performing these tests, the
following values of resistor R may be used:
Type CF — 100 kOhm (220 to 240 V)
Disconnect safety tester from line voltage before connecting
patient cable/test plug to device.
After line voltage has been disconnected, connect the
patient cable/test plug to device.
This test performs leakage current tests during single fault conditions
(S.F.C.) with line voltage applied to the floating (patient connection)
inputs.
N.C.
S.F.C
Polarity: NORM &
RVS S1
S1 (neutral): closed
GND: closed
Polarity: NORM &
RVS S1 (neutral): closed I
open GND: open I
Total tests/combinations: 2
Total tests/combinations: 4
10 uA (IEC & UL)
50 uA (EC & UL)
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...