Maintenance:
Preparing System for FRU Replacement
Required Tools and Manuals
PWA Connector Details
The legends for the PWA Connectors (J1, J2 etc) for the new PWA FRU
2066447-002 are different as compared with the old PWA FRU 2047330-
001. Refer the following table for the old and new connector legends:
Table 4-1. Required tools and manuals
Sl no:
Description
1
Phillips screwdriver
2
Anti-static wrist strap
3
Service manual as needed for reference
4
Safety tester for measurements according to 60601-1
5
Connecting wires and clips required for electrical
safety tests (optionl)
6
Multi-meter
7
Multi Lead ECG simulator
8
MAC 600 Operator’s Manual for reference
9
Nose Plier
10
Recommended SD card
Table 4-2. PWA Connectors
Connector Description
OLD PWA FRU
PN: 2047330-001
NEW PWA FRU
PN:2066447-002
Key sheet
J4
J7
Display
J5
J6
Printer
J10
J5
SD Card
J1
J3
Power In
W1
W1
RS232
J2
J2
JATAG
J3
J8
USB
J9
J1
Modem
J12
J9
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...