Revision G
MAC
TM
600
3-11
2047426-002
Troubleshooting:
General Fault Isolation
Writer Test
1. From the
Diagnostics menu
, select
Writer Tests
.
The following window opens.
2. Perform the
50mm/s Speed
Test.
a. Select the
50mm/s Speed Test
button. The writer prints the 50
mm/s speed test report.
b. Examine the printed report. Use the following criteria to
determine if the writer passed or failed the 50mm/s speed test.
Count the number of subdivisions printed in a 20mm grid.
If the count is between 19-21, the test has passed.
NOTE
We recommend that you use a wider grid such as 50mm for
more accurate readings. For a 50mm span, the expected
result is a count of 45-55.
3. Repeat step 2 for the 25mm/s and 5mm/s speed tests. The pass-fail
criteria for each of the remaining tests are the same as indicated
above.
4. Perform the
Roller Test
.
a. Check that the test pattern printing is consistent. A white or
black line across the pattern indicates a defective or missing
print head dot. Replace the thermal printer assembly. See
“Replacing the Thermal Printer”
5. Perform the
Continuously Run Out Paper Test
.
This test continuously rolls out the paper. It also shows the number
of pages ejected if the unit is set in Z-fold paper mode.
If the number of pages does not increase while in Z-fold paper mode,
and if you run out of paper, there is a problem with the Thermal
Printer Cue sensor or the paper is not placed correctly. Check the
paper placement, if the problem persists replace thermal printer
assembly. See
“Replacing the Thermal Printer”
6. Select
Main menu
and press
Enter
to return to
System Diagnostics.
Содержание MAC 600 SPY
Страница 1: ...2047426 002 Revision G GE Healthcare MAC 600 Service Manual Product Code SPY ...
Страница 7: ...Revision G MACTM 600 1 1 2047426 002 1 Introduction ...
Страница 15: ...Revision G MACTM 600 1 9 2047426 002 Introduction Service Information For your notes ...
Страница 16: ...1 10 MACTM 600 Revision G 2047426 002 Introduction Service Information ...
Страница 17: ...Revision G MACTM 600 2 1 2047426 002 2 Equipment Overview ...
Страница 25: ...Revision G MACTM 600 3 1 2047426 002 3 Troubleshooting ...
Страница 26: ...3 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 41: ...Revision G MACTM 600 3 17 2047426 002 Troubleshooting Equipment Problems Paper Moving But Not Printing ...
Страница 42: ...3 18 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Printer Door Open Message ...
Страница 43: ...Revision G MACTM 600 3 19 2047426 002 Troubleshooting Equipment Problems Printing Only Baselines ...
Страница 44: ...3 20 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems Battery Error Message ...
Страница 45: ...Revision G MACTM 600 3 21 2047426 002 Troubleshooting Equipment Problems Unit Not Turning ON in Battery Mode ...
Страница 46: ...3 22 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems External Power LED Not Lighting Up ...
Страница 50: ...3 26 MACTM 600 Revision G 2047426 002 Troubleshooting Equipment Problems ...
Страница 51: ...Revision G MACTM 600 4 1 2047426 002 4 Maintenance ...
Страница 52: ...4 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 85: ...Revision G MACTM 600 5 1 2047426 002 5 Parts and Replacements ...
Страница 86: ...5 2 MACTM 600 Revision G 2047426 002 For your notes ...
Страница 110: ...5 26 MACTM 600 Revision G 2047426 002 Parts and Replacements Ordering Parts ...
Страница 111: ...Revision G MACTM 600 A 1 2047426 002 A Technical Specification ...
Страница 116: ...A 6 MACTM 600 Revision G 2047426 002 Technical Specification ...
Страница 117: ...Revision G MACTM 600 B 1 2047426 002 B Electromagnetic Compatibility ...
Страница 123: ...Revision G MACTM 600 i 1 2047426 002 Index ...
Страница 126: ...Index i 4 MACTM 600 Revision G 2047426 002 ...