MAC™ 5500

Resting ECG Analysis System

Service Manual

2020299-020         Revision E

GE Healthcare

Содержание MAC 5500

Страница 1: ...g MAC 5500 Resting ECG Analysis System Service Manual 2020299 020 Revision E GE Healthcare ...

Страница 2: ...y to earlier software versions Due to continuing product innovation specifications in this manual are subject to change without notice MUSE CASE MAC MARS and 12SL are trademarks owned by GE Medical Systems Information Technologies a General Electric Company going to market as GE Healthcare All other trademarks are owned by their respective owners 2005 2010 General Electric Company All rights reser...

Страница 3: ... Information 1 8 Service Requirements 1 8 Equipment Identification 1 8 Serial Number Format 1 9 Label Format 1 9 General Description 2 3 Front View 2 3 Back View 2 4 Connector Identification 2 5 MAC 5500 Block Diagram 2 6 Theory of Operation 2 7 Overview Block Diagram 2 7 Power Supplies 2 10 3V C 2 10 3V M 2 10 3V EMI 2 10 5V M 2 10 5V EMI 2 10 18V 2 11 1 8V 2 11 12V 2 11 REF2V5 2 11 VAna VAna 2 1...

Страница 4: ...LCD Panel EMI Reduction Components 2 26 CRT Video DAC Sync Buffers 2 26 Acquisition Module Transceiver Power Switch 2 27 Transceiver 2 27 Acquisition Power Regulator Switch 2 27 COMM Port Power Switch Current Limiter KISS Power 2 27 Thermal Printhead Power Pixel Test Hardware 2 28 Super I O Peripheral Controller 2 28 RS 232 Serial Ports One Dual Mode RS 232 IrDA 2 28 Clock Calendar 2 29 PS2 Keyboa...

Страница 5: ...ter Password 4 8 System Diagnostics Main Menu 4 9 Display Tests 4 9 Pixel Verification Test 4 9 Grey Scale Test Patterns 4 9 Speaker Test 4 10 Keyboard Test 4 10 Writer Tests 4 11 Battery Tests 4 13 Battery Status 4 13 Battery Discharge Test 4 13 Battery Charge Test 4 14 Print Charge Discharge Test Results 4 14 Communication Tests 4 14 COM Port Loopback Test 4 14 External Modem Test 4 15 Internal ...

Страница 6: ...Editing 4 22 Entering Patient Data 4 23 Transmission 4 23 Losing Fields When Transmitting 4 23 Input and Output Connectors 4 24 A Pins J1 4 24 COM1 COM3 4 Pins J3 4 24 COM2 Pins J5 4 25 Analog Pins J6 4 25 EXT VID Pins J7 4 26 CPU PCB Input Output Signals 4 27 Battery Pack Monitor J2 4 27 LCD Backlight J4 4 27 Keyboard J8 4 27 LCD J10 4 28 Power Supply Motor J11 4 29 Thermal Printer J12 4 30 Flopp...

Страница 7: ...eassembly 5 12 Reassembly 5 12 Main CPU Board 5 13 Removal of CPU Board 5 13 Reassembly of CPU Board 5 13 Software 5 15 Service Only Setups 5 16 Restore System Setups 5 17 Restore Options 5 17 Disable Options 5 18 Printhead Replacement 5 19 Removal 5 19 Reassembly 5 19 COMM Board Replacement 5 19 Setting up LAN Communications 5 20 Writer Roller Carriage Assembly 5 22 Removal 5 22 Reassembly 5 22 T...

Страница 8: ... 4 Vectorcardiography B 4 Hi Res and PHi Res Signal Averaged Electrocardiography B 5 Physical B 5 Environmental B 5 Safety B 7 Appendix C Electromagnetic Compatibility Electromagnetic Compatibility EMC C 3 Guidance and Manufacturer s Declaration Electromagnetic Emissions C 3 Guidance and Manufacturer s Declaration Electromagnetic Immunity C 4 Guidance and Manufacturer s Declaration Electromagnetic...

Страница 9: ...Revision E MAC 5500 resting ECG analysis system 1 1 2020299 020 Introduction 1 Introduction ...

Страница 10: ...1 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction For your notes ...

Страница 11: ...red field repairable Where necessary the manual identifies additional sources of relevant information and or technical assistance See the operator s manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use Intended Audience This manual is intended for the person who uses maintains or troubleshoots this equipment Table 1 Revision History...

Страница 12: ...rd is defined as a source of potential injury to a person Term Definition DANGER Indicates an imminent hazard which if not avoided will result in death or serious injury WARNING Indicates a potential hazard or unsafe practice which if not avoided could result in death or serious injury CAUTION Indicates a potential hazard or unsafe practice which if not avoided could result in minor personal injur...

Страница 13: ...th conductor is in doubt CAUTIONS This equipment contains no serviceable parts Refer servicing to qualified service personnel U S Federal law restricts this device to the sale by or on the order of a physician Responsibility of the Manufacturer GE Medical Systems Information Technologies is responsible for the effects of safety reliability and performance only if Assembly operations extensions rea...

Страница 14: ...recommended by GE Medical Systems Information Technologies Contact GE Medical Systems Information Technologies for information before connecting any devices to this equipment that are not recommended in this manual If the installation of this equipment in the USA will use 240 V rather than 120 V the source must be a center tapped 240 V single phase circuit Parts and accessories used must meet the ...

Страница 15: ...is system 1 7 2020299 020 Introduction Equipment Symbols Equipment Symbols For a list and examples of the symbols that appear on the equipment and packaging refer to the MAC 5500 Resting ECG Analysis System Operator s Manual PN 2020299 153 ...

Страница 16: ...er warranty voids that warranty It is the user s responsibility to report the need for service to GE or to one of their authorized agents Equipment Identification The serial number label is located inside the device where shown below Every GE Medical Systems Information Technologies device has a unique serial number for identification The serial number is formatted as shown in Serial Number Format...

Страница 17: ...MAC 5500 with product code SCD B Year manufactured 00 99 00 2000 01 2001 02 2002 and so on C Fiscal week manufactured D Production sequence number E Manufacturing site F Miscellaneous characteristic A B C D E F Table 3 Equipment Identification Label A Date of manufacture in YYYY MM format B Part number of product C Product code description D Serial number described above E Manufacturing site A B C...

Страница 18: ...1 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Introduction Service Information ...

Страница 19: ...Revision E MAC 5500 resting ECG analysis system 2 1 2020299 020 Equipment Overview 2 Equipment Overview ...

Страница 20: ...2 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview For your notes ...

Страница 21: ...ctrocardiography There are also options for communication capabilities Front View Name Description A display screen View the waveform and text data B modem port Connect the telephone cable here C LAN port Connect to the LAN here The green LED right of this port indicates that power is supplied to the communication card from the Ethernet link The amber LED left of this port flashes to indicate netw...

Страница 22: ...el connectors Connect peripheral devices here B secure data card slot Insert secure data card for external storage here C green AC power light Indicates the system is connected to AC power D amber battery light Indicates the battery is recharging E internal access button Press to open the system to change paper or the battery A B D E C 117A ...

Страница 23: ... the stress option connect a T2000 or external blood pressure device cable to this port C 2 Connect a local transmission cable serial line modem or client bridge wireless option D ANA TTL Connect a device requiring analog data or TTL trigger ultrasound stress echo ergometer analog treadmill blood pressure units etc E EXT VID Connect an external video display F IR Point at a MAC 5000 MAC 5500 or MU...

Страница 24: ...system Revision E 2020299 020 Equipment Overview MAC 5500 Block Diagram MAC 5500 Block Diagram Following is the basic diagram of the MAC 5500 with the 007 board For additional information refer to Overview Block Diagram on page 2 7 79A ...

Страница 25: ...olt Power Switch Acquisition Module Transceiver Power Switch Printhead Power Switches and Pixel Test Circuit Daughter Board Interface which support LAN and modem Switch Mode Power Supplies 3 3 Volt for Logic LCD 5 Volt for Logic Printer 12 Volt for LCD backlight External Com Port Power Battery Charger 12 Charge Pump for Analog Circuits Linear Power Supplies 1 8 Volt AT91RM9200 Core and FPGA Core 2...

Страница 26: ... bit PIO Controllers with Up to 122 Programmable I O Lines Input Change Interrupt and Open drain Capability on Each Line 20 channel Peripheral Data Controller DMA Multimedia Card Interface MCI Automatic Protocol Control and Fast Automatic Data Transfers MMC and SD Memory Card compliant Supports Up to Two SD Memory Cards Three Synchronous Serial Controllers SSC Independent Clock and Frame Sync Sign...

Страница 27: ...ace Thermal Printhead Interface Serial EEPROM Interface BBus Controller Four PWM Analog Outputs Beep Generator PC Card Interface A PC Super I O controller containing Two Serial Ports one dual mode RS 232 IrDA Clock Calendar Y2K compliant PS 2 Keyboard Port for card and barcode readers Floppy Disk Controller Three Peripheral Microcontrollers The Three Stooges System Supervisor Battery Charger Gauge...

Страница 28: ...be used sparingly NOTE The MAX782 s low current regulator is dreadfully inefficient Regulator Q current appears to be about 3x the load current This makes conservation of load on 3V C crucial 3V M Most of the MAC 5500 hardware runs from 3V M The MAX782 provides this rail from the battery via a PWM synchronous switching regulator Moe controls 3V M in tandem with 5V M 3V EMI This is simply an RF blo...

Страница 29: ...put cannot provide sufficient current for all of the systems 12V loads so a secondary 12V regulator is required The Main 12V Regulator U17 a switching buck regulator provides the higher currents needed by these loads A P channel MOSFET Q1 switch precedes the regulator to provide on off control Gate capacitor C12 slows the turn on off time of the MOSFET switch to eliminate switching transients The ...

Страница 30: ...The FPGA provides many functions including the acquisition interface the printer interface and the Stooges interface Bbus to name a few The FPGA uses a built in frequency doubler to raise this 24 Mhz clock to 48 Mhz for internal use All functions inside the FPGA use the clocks derived from 48MHz The main derived clocks are 1 MHz for acquisition interface 4 MHz for printer data shift clock interfac...

Страница 31: ...nal LCD controller with a separate video memory interface compensates overall performance of the CPU External Bus Interface The external bus interface width is limited to 16bit in ATMEL CPU when compare with the 32 bit interface of StrongARM All the non VGA FPGA registers are either 8 bit or 16 bit wide However all these were accessed using 32 bit access in 005 board and aligned to 32 bit word To ...

Страница 32: ...esent at the Xbus while toggling CCLK This is achieved by toggling the NAND_RE alternately with CCLK The NAND_RE needs to be under the GPIO control instead of static memory controller to do this The ALE and CLE are also controlled in GPIO mode and tied to low level during read cycle while configuring the FPGA The CLE and ALE acts as address line A23 and A25 respectively during Address and command ...

Страница 33: ...y The service menu provides a provision to update the SPI data flash with the primary boot program copy residing the NAND Flash FPGA Internal Logic All of the MAC 5500 resting ECG analysis system s proprietary hardware is contained in a single Xilinx FPGA that contains XBus Controller Video Interface LCD Controller with SDRAM frame buffer Video Waveform Scroller Interrupt Controller System Interru...

Страница 34: ...r I O controller and NAND Flash ICs are both located on this bus Unlike the 3 3V only main data address busses XBus is compatible with both 5V and 3 3V logic To maintain software compatibility with previous board versions the low order address byte is not used by XBus Starting XBus addressing with A8 also produces Super I O addresses that easily map to their standard PC equivalents simply append 0...

Страница 35: ...trol register in the SDRAM and an initial burst of refresh cycles are performed to prepare the memory array for operations The SDRAM controller does this all automatically at startup requiring no initialization by the CPU SDRAMs being dynamic require periodic refresh to maintain the contents of the memory array The SDRAM controller performs this refresh automatically between accesses All details o...

Страница 36: ... and a dynamic pixel byte each in their own lanes When the CPU writes a pixel to the static plane the upper five bits of the byte are routed to the static byte lane the lower three bits are set to zero and the dynamic byte lane is disabled When the CPU reads a static pixel both the static and dynamic byte lanes are accessed but only the upper five bits of the static byte lane are passed on to the ...

Страница 37: ...re required to fill large regions of the display To reduce both CPU and frame buffer loading the LCD controller provides a simple fill engine which automates the filling of rectangular regions of the frame buffer and takes advantage of the burst capabilities of the SDRAM The fill engine interface is simple consisting of four boundary registers to define the fill region and one register to record t...

Страница 38: ...od and to synchronize display related processes in the CPU such as waveform drawing and scrolling control A similar interrupt is provided to signal the completion of fills Both interrupts may be disabled and or acknowledged in the system control registers Video Waveform Scroller There are numerous ways of achieving a scrolling waveform none of which is supported by standard LCD controllers The MAC...

Страница 39: ...is added to the video signal The video sync signal may be disabled under software control to accommodate monitors that do not accept sync on green The sync signal is applied to all three video guns to eliminate color shifting in systems that do not perform blank level video clamping TTL Sync For monitors that do not accept sync on green TTL logic level horizontal and vertical sync signals are prov...

Страница 40: ... the acquisition module Details A constant reference clock frequency of 1MHz must be provided to the acquisition module for generation of its internal sampling clocks To eliminate the need for data lines command information is encoded on this reference clock by altering its duty cycle The FPGA provides a serializer for the command bytes and clock generator modulator to transmit both the clock and ...

Страница 41: ...age is critically low Each print line requires 1728 bits of data To conserve FPGA resources each line is divided into three chunks of 512 bits each with one leftover chunk of 192 bits The FPGA provides a single 32 word x 16 bit buffer 512 bits to hold the print line data After writing a chunk of data to the buffer the ATMEL CPU enables serialization of the data by reading one of two registers to s...

Страница 42: ...ended for ECG output and produces a swing of 10 to 10V The two 8 bit channels provide a unipolar 10V output Regardless of the resolution or swing range of each PWM channel the FPGA treats the data value as a signed 16 bit number representing a voltage from 10V 0x7fff to 10V 0x8000 Logic in each PWM channel ensures that the closest possible voltage is generated for each data value ex 0x8000 on an 8...

Страница 43: ...CPU clock or 59 904 Mhz Though the size requirement is less the video frame buffer also use 256Mbit SDRAM NAND Flash There are two 32 Mbytes NAND Flash in 006 and 007 boards One is used for storing FPGA configuration data and system software The other is for data storage The access to NAND flash is through a dedicated smart media interface logic provided by ATMEL CPU Unlike the 005 board NAND flas...

Страница 44: ... external hardware is required to generate the analog video levels expected by external VGA monitors LCD Panel EMI Reduction Components To reduce EMI 47pF capacitors have been added to all LCD digital lines In addition 49 9 resistors have been added to the video clock and Sync lines CRT Video DAC Sync Buffers A triple 6 bit video DAC supports external analog VGA monitors Only one DAC Level Shifter...

Страница 45: ...The regulator also has build in current limit and over temperature shutdown for protection COMM Port Power Switch Current Limiter KISS Power Power for external peripherals such as a modem or the KISS vacuum electrode pump is available on the COMM connectors Power may be turned on off under software control and current limiting is employed to protect internal operations from excessive external load...

Страница 46: ... four analog inputs he continuously monitors By loading a single black dot into the print head it is possible to measure its resistance A typical TPH has an average dot resistance of 650 Presuming negligible driver leakage current a single enabled dot would drop 3 9V While there are mitigating influences off pixel driver leakage current and on pixel driver saturation voltage that might make accura...

Страница 47: ... functions are self contained and reasonably portable Originally there were four Stooges Since the boot loading and FPGA configuration is handled by ATMEL CPU the Curly was removed in 006 board Although there are three of these little fellows in the MAC 5500 each performing a different function there is only one firmware image By merging the code from each of the three functions into a single ROM ...

Страница 48: ...of treadmill control keys up down faster slower A special key code indicates when all keys are up as a safeguard against stuck keys in the application software Unlike previous keyboard encoder designs Shemp does not provide dedicated scan hardware for the shift and or option keys These keys are now located in the scan matrix Careful placement of keys in the scan matrix allows simultaneous depressi...

Страница 49: ...ve torque has become excessive or the motor has stalled This condition is reported as a Paper Jam Error Similarly if the servo error variable saturates at off for more than a predetermined time it is assumed that the someone is pulling on the paper with a force that exceeds the paper drive system torque and as a result paper speed has been pulled out of regulation This condition is reported as a P...

Страница 50: ...t When SYSRESET is released ATMEL CPU configures the FPGA and load secondary boot program from information stored in NAND Flash Moe expects the ATMEL CPU to request status via the BBus interface after startup If that request doesn t arrive in time Moe places the system back in reset and removes power The time is set as two minutes six seconds for this version This time out considers the time for s...

Страница 51: ...ormal system operating power are obtained from the AC power supply The charger circuitry monitors both battery charge current and power supply output current The battery is always charged at the maximum rate possible but system power demands take precedence over charger demands The charger automatically reduces charge current as required to keep the AC power supply output current within specified ...

Страница 52: ...ve a false full reading on the gas gauge The battery and ambient thermistors are the same type and value to ensure reasonable tracking Capacitors C155 and C156 filter noise from the temperature sense lines Thermistor Bias Switch To reduce quiescent power consumption when the system is turned off a switch disables bias current to the battery and ambient thermistors Q104 under control of MOE switche...

Страница 53: ...ery pack s state of charge Moe s A D converter hasn t sufficient dynamic range to cover the full range of system currents at high resolution so some compromises must be made The current monitor s full scale range is set to a value that is likely to encompass normal operating currents Peaks above this level 6Amps are clipped The effects of this clipping are minimal as such high density printing occ...

Страница 54: ...2 36 MAC 5500 resting ECG analysis system Revision E 2020299 020 Equipment Overview Theory of Operation ...

Страница 55: ...Revision E MAC 5500 resting ECG analysis system 3 1 2020299 020 Installation 3 Installation ...

Страница 56: ...3 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation For your notes ...

Страница 57: ...tional MAC 5500 Trolley Use this picture for reference when installing trolley options NOTE Because the optional Trolley is made by another vendor for GE Medical Systems Information Technologies the serial number format is different from that shown in Chapter 1 MAC 5500 Acquisition Module Arm and Holder Front Cover Locking Casters Swivel Casters Trolley Serial 13A ...

Страница 58: ...45 cm 33 25 inches The trolley is normally shipped at the 92 07 cm 36 25 inches height but can be changed to fit your needs To change to the lower height use the following steps 1 Tip the trolley on its side and using a 1 2 inch socket remove the 4 outer 1 2 inch bolts and slide the base assemble up on the column 2 Remove the remaining bolts and mounting plate 3 Flip the mounting plate and reverse...

Страница 59: ...Revision E MAC 5500 resting ECG analysis system 3 5 2020299 020 Installation Preparation for Use CAUTION Do not over tighten Over tightening the bolts may cause them to strip 17A 16A ...

Страница 60: ... analysis system To secure the MAC 5500 to the trolley assembly follow these steps 1 Lock the wheels to prevent the trolley from rolling 2 Remove the end panel by pulling out and up 3 Place the unit on the trolley surface then slide it on until the unit is firmly in place and under the tab at the rear of the on the tray 18A 19A 20A ...

Страница 61: ...llation Preparation for Use 4 Secure the MAC 5500 to the trolley by tightening the three captive screws located under the trolley tray 5 Replace the end panel by pushing up and in until you hear a snap 6 Unlock the wheels to allow free movement of the trolley 21A 22A 23A ...

Страница 62: ...comes assembled and ready to install on the trolley To install a modem kit on the trolley complete the following steps 1 Find the modem mounting site located under the Acquisition Module support arm at the rear of the trolley where the kit is to be installed 2 Slide the assembly up in place so that the bracket slot catches on the bracket lip 3 Tighten the three mounting screws to secure the modem ...

Страница 63: ... 5500 resting ECG analysis system 3 9 2020299 020 Installation Preparation for Use 4 Plug the modem cable into connector port 2 on the MAC 5500 5 Refer to the operator s manual for information on using the modem 30A ...

Страница 64: ...two different trolley styles Disregard and do not use the parts indicated in the following illustration To install the Magnetic Card Reader and its mounting bracket on the trolley complete the following steps 1 Remove both end panels by pulling out and up at the bottom 2 Using a Phillips screw driver fasten the card reader assembly under the front handle Align with holes provided under front handl...

Страница 65: ...ley column towards the rear as shown below 4 At the front hold the cable to the side so it clears the front panel as you replace the panel 5 Plug the cable connector into port A then replace the back panel 6 Refer to the MAC 5500 Operator s Manual for information on using the Magnetic Card Reader Cable Routing Rear View 34A 77A 70A ...

Страница 66: ...l the Bar Code Reader and its cable mounting bracket on the trolley complete the following steps 1 Fasten the cable clamp bracket to the underside of the rear handle using a Phillips screw driver and the self tapping screws provided NOTE DO NOT overtighten Overtightening the screw may cause the screw to strip and clamp to fail Cable Clamp Cable Clamp Bracket Barcode Reader 71A 72A Cable Clamp Cabl...

Страница 67: ...o place the correct amount of slack to free the cable from stress when the MAC 5500 needs to be re opened 3 Next fasten the cable and clamp to the clamp bracket then close the MAC 5500 Observe that there is enough slack to allow free movement of the cable when re opening the MAC 5500 4 Refer to the MAC 5500 Operator s Manual for information on how to use the Bar Code Reader Port A Internal Access ...

Страница 68: ...n E 2020299 020 Installation Preparation for Use Type S Trolley Assembly 1 To mount the MAC 5500 to the Type S trolley follow the steps in the illustration below 2 Route patient cable through trolley and fasten with cable clamp as shown below 75A 76A ...

Страница 69: ...e Main Menu complete the following in the order shown below Select System Setup Enter System password Exercise Test Inputs Outputs Change Blood Pressure to Nipon Colin Sun Tech Model Tango Connection Requirements Use cable PN 2008113 001 to connect from the MAC 5500 port 1 to the Sun Tech serial port Use cable PN 2008111 001 to connect from the MAC5500 ANA TTL port to the Sun Tech QRS trigger inpu...

Страница 70: ...n Technologies Treadmills Model T2000 Connection Requirements Use cable PN 2007918 001 T2000 to connect from the MAC 5500 port 1 to the treadmill serial port Device Configuration Requirements None MAC 5500 Configuration Requirements Use the Edit Protocol application to set the protocol Test Type to Treadmill in MPH or Treadmill in Km H for protocols that will be used with this treadmill Analog Tre...

Страница 71: ...81 001 Lode Ergometer to connect from the MAC 5500 ANA TTL port to the ergometer analog control port NOTE For any other ergometer the customer is responsible for making the appropriate cable Device Configuration Requirements Refer to ergometer Operator s Manual MAC 5500 Configuration Requirements Use the Edit Protocol application to set the protocol Test Type to Ergometer in Watts or Ergometer in ...

Страница 72: ...3 18 MAC 5500 resting ECG analysis system Revision E 2020299 020 Installation MAC 5500 ST Requirements and Configuration For your notes ...

Страница 73: ...Revision E MAC 5500 resting ECG analysis system 4 1 2020299 020 Troubleshooting 4 Troubleshooting ...

Страница 74: ...4 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Troubleshooting For your notes ...

Страница 75: ...riptions Introduction The troubleshooting information in this chapter helps you narrow service problems to one of the replaceable assemblies These assemblies illustrated in the block diagram are discussed in more detail in the individual assembly chapters along with replacement procedures Assembly Block Diagram 121A ...

Страница 76: ...ays If the equipment is not working properly ask yourself the following questions Is the unit turned on Have there been any changes in the use location or environment of the equipment that could cause the failure Has the equipment hardware or software been modified since last use Is operator error the cause of the problem Try to repeat the scenario exactly and compare that to the proper operation ...

Страница 77: ...Revision E MAC 5500 resting ECG analysis system 4 5 2020299 020 Troubleshooting General Fault Isolation Power up Flow Chart 60B ...

Страница 78: ...n Poor Quality ECGs Poor quality ECGs can be caused by factors in the environment inadequate patient preparation hardware failures related to the acquisition module leadwires cables or problems in the unit Use a simulator to obtain an ECG report If the report is good the problem is external to the unit ...

Страница 79: ...ing Loose screws in plugs Fuses Type and rating Replace as necessary Interface Cables Excessive tension or wear Loose connection Strain reliefs out of place Circuit Boards Moisture dust or debris top and bottom Loose or missing components Burn damage or smell of over heated components Socketed components not firmly seated PCB not seated properly in edge connectors Solder problems cracks splashes o...

Страница 80: ...te Master Password later in this section 5 When the System Setup menu displays hold down Shift and press F5 Shift F5 6 Type prod at the service password prompt 7 The System Diagnostics menu appears Substitute Master Password If you do not have access to the system s password you can create a master password as follows 1 At the prompt for the system password enter meimac A random 6 digit number dis...

Страница 81: ...ay Tests to verify that all the screen pixels are working and that the brightness and contrast samples appear to be within normal range There are no screen display adjustments The screen Display Tests are as follows Pixel Verification Test Use the arrow pad control to move the bar across the screen and look for any missing pixels on the display Press the F1 key to turn on all of the pixels simulta...

Страница 82: ...creen and also displayed at the top of the screen It is normal for a dim background image to remain on the screen when you select the next key The numeric value that displays at the top of the screen is the scan code representation of the pressed key NOTE The display shows keys in the upper part of the screen that are only available with the MAC 5500 ST keyboard Check both of the Shift keys by pre...

Страница 83: ...agonal lines printed across the pages should be straight and uniform without curves or wavering The perforations should align with the tear bar on the door after cueing Paper travel should be smooth C Scan Tests 1 2 3 These tests are combinations of test pattern I and the roller test They are used by the writer vendor 50 mm s Test Pattern I 25 mm s Test Pattern I and 5 mm s Test Pattern I These te...

Страница 84: ...oller and may indicate that the print carriage assembly needs to be replaced A white line across the length of the page indicates a missing print head dot Test Pattern II A combination of Test Pattern I and Roller tests The first three pages consist of a series of triangular waveforms and various hashmarks The fourth page is a partial roller test Test Pattern II Continuous Test Pattern II runs con...

Страница 85: ...nt temperature inside the unit If the ambient temperature inside the unit is more than 10 greater than the current room temperature Maximum and minimum ambient temperature Current battery charging status Battery Discharge Test This test charges the battery to full capacity if necessary then monitors a discharge cycle NOTE To cancel the test at anytime press the ESC key 1 To perform the battery dis...

Страница 86: ...n to charge 4 When the battery charge test is complete go to Print the Charge Test Results Monitored information written to internal memory includes Charge rate in mAH Battery temperature Battery charge status Percent of charge Print Charge Discharge Test Results Prints the results of the last discharge or charge test to the writer Communication Tests COM Port Loopback Test The Communications Port...

Страница 87: ... the Ethernet module test is unsuccessful the ID and firmware rev display N A Press any key to exit this test Acquisition Module Test Follow the instructions on screen NOTE A shorting bar is required to perform this test Tests if the acquisition module is powered Tests if the acquisition module is communicating Displays the acquisition module lead wire noise Indicates when one of the three acquisi...

Страница 88: ...firms that the correct values are read from the Analog Inputs A pass fail result displays If either test fails the main CPU board may need to be replaced NOTE Pin numbers refer to the ANA TTL port Floppy Drive Tests This test does not apply to the MAC 5500 system Follow the instructions on screen A read write test is performed on a formatted floppy disk and a pass fail test result is displayed Try...

Страница 89: ...ormat the internal storage memory if it contains data which has not yet been transferred SD Card Tests 1 Insert SD card which is not write protected 2 Select SD Card Tests 3 If test fails replace SD card and repeat test 4 If test still fails CPU board may need to be replaced NOTE Do not format the SD card if it contains data which has not yet been transferred ...

Страница 90: ...ng the acquisition of a resting ECG Missing ACI TIPI Report ACI TIPI is disabled Enable ACI TIPI The selected report is without interpretation Select Interpretation for the report The ACI TIPI required information is not entered Make sure the age range gender and chest pain complaints are entered The patient was entered as pediatric Make sure you enter an age range less than 16 The original ECG wa...

Страница 91: ...charge is low Connect the system to an AC wall outlet to charge the battery appears on the screen The writer door is open Close the writer door The system does not power up when operating from battery power The battery is empty Connect the system to an AC wall outlet to charge the battery The system shuts down when operating from battery power Battery is empty or the Automatic Shutdown feature is ...

Страница 92: ... the SD card is not write protected Try a new SD card If your system is not set up to automatically save records you must manually save by pressing Store Format an SD Card Q How do I format an SD card in the MAC 5500 A Most secure digital cards do not require formatting In the event an unformatted SD card is used with the system the following message will display This SD Card cannot be read and re...

Страница 93: ...etup to save you from entering it for each test Go to System Setup Select Basic System Select Miscellaneous Setup Arrow down to Location and type in the number you want set as your default Then press Enter Press the esc key until you return to System Setup Select Save Setup Select To System You can select Main Menu to exit System Setup Patient Questions Q How do I change what questions I see when ...

Страница 94: ...e number of copies you want in the with column If you do not want the MAC 5500 interpretation to print on the ECG put the number of copies you want in the without column Click view report type to see the examples of the report formats Press the esc key until you return to System Setup Select Save Setup Select To System You can select Main Menu to exit System Setup Hi Res and Phi Res Q What is the ...

Страница 95: ...umber at the least If you are transmitting to the MUSE system you will want to enter the Location number as well If an emergency situation dictates that you must complete the test without entering the patient data make sure you edit the record to add the missing information before you transmit it to the MUSE system Transmission Losing Fields When Transmitting Q Why do I lose the Referring MD and T...

Страница 96: ...r those connectors The pin by pin descriptions identify the signal names and pin outs for each connector on the unit A Pins J1 COM1 COM3 4 Pins J3 Table 2 A Pins J1 Pin Name 1 Data 2 NC 3 Ground 4 5V 5 Clock 6 NC Table 3 COM1 COM3 4 Pins J3 Pin COM1 Signal COM3 4 Signal 1 RTS COM3 TxD 2 CTS COM3 RxD 3 TxD 4 Ground 5 RxD 6 DTR COM4 TxD 7 12V 8 DSR COM4 RxD 2 1 6 5 3 4 66A 6 3 1 8 5 4 2 7 67A ...

Страница 97: ...tors COM2 Pins J5 Analog Pins J6 Table 4 COM2 Pins J5 Pin Name 1 RTS 2 CTS 3 TxD 4 Ground 5 RxD 6 DTR 7 12V 8 DSR 6 3 1 8 5 4 2 7 67A Table 5 Acquisition Module Connector J6 Pin Name 1 12V 2 DC Output 1 3 TTL Trigger Output 4 Ground 5 Ground 6 DC Output 2 7 DC Input 1 8 ECG Output 9 DC Input 2 6 9 5 1 68A ...

Страница 98: ...bleshooting Input and Output Connectors EXT VID Pins J7 Table 6 External VGA Video J7 Pin Name 1 Red Video 2 Green Video 3 Blue Video 4 Ground 5 Ground 6 Ground 7 Ground 8 Ground 9 NC 10 Ground 11 Ground 12 NC 13 Horizontal Sync 14 Vertical Sync 15 NC 1 6 15 5 11 10 68A ...

Страница 99: ...Signal 1 18V Battery Power 2 18V Battery Power 3 Battery Temperature Sense 4 3V Temperature Sense Power 5 Battery Ground 6 Battery Ground LCD Backlight J4 Pin No Signal 1 12V Power 2 12V Power 3 12V Power 4 Ground 5 Ground 6 Brightness Select 7 Backlight Enable 8 NC 9 Ground 10 Ground Keyboard J8 Pin No Signal 1 NC 2 NC 3 NC 4 NC 5 NC 6 Sense4 7 Sense2 8 Sense1 9 Sense0 ...

Страница 100: ...se5 12 Sense6 13 Sense7 14 Drive0 15 Drive1 16 Drive2 17 Drive3 18 Drive4 19 Ground 20 Power Key 21 Drive5 22 Drive6 23 Drive7 24 Drive8 25 Drive9 26 Drive10 LCD J10 Pin No Signal 1 Ground 2 Pixel Clock 3 Hsync 4 Vsync 5 Ground 6 R0 LSB 7 R1 8 R2 9 R3 10 R4 11 R5 MSB 12 Ground 13 G0 LSB 14 G1 15 G2 16 G3 Keyboard J8 Continued ...

Страница 101: ... 20 B0 LSB 21 B1 22 B2 23 B3 24 B4 25 B5 MSB 26 Ground 27 Data Enable 28 3V Power 29 3V Power 30 NC 31 NC Power Supply Motor J11 Pin No Signal 1 Motor Encoder B 2 5V Power 3 Motor A 4 Motor Encoder A 5 Ground 6 Motor B 7 NC 8 28V Power 9 Ground 10 Battery Charge LED 11 28V Power 12 Ground 13 Door Open Detect 14 Ground LCD J10 Continued ...

Страница 102: ...al Printer Power 3 Thermal Printer Power 4 Thermal Printer Power 5 Thermal Printer Power 6 Thermal Printer Power 7 Thermal Printer Power 8 Ground 9 Ground 10 Ground 11 Ground 12 Ground 13 Ground 14 Ground 15 Cue Sense 16 NC 17 5V Main Power 18 Ground 19 Data Strobe 20 Data Strobe 21 Data Strobe 22 Data Strobe 23 Data Load 24 Data Clock 25 Print Head Temperature 26 Pixel Data ...

Страница 103: ...ppy drive not installed Pin No Signal 1 5V Power 2 Index 3 5V Power 4 Drive Select 0 5 5V Power 6 Disk Change 7 NC 8 Media Sense 0 9 Media Sense 1 10 Motor Select 0 11 NC 12 Direction 13 NC 14 Step 15 Ground 16 Write Data 17 Ground 18 Write Gate 19 Ground 20 Track 0 21 Ground 22 Write Protect 23 Ground 24 Read Data 25 Ground 26 Head Select ...

Страница 104: ... resting ECG analysis system Revision E 2020299 020 Troubleshooting CPU PCB Input Output Signals Acquisition Module J14 Pin No Signal 1 Power 2 Ground 3 TX RS485 4 TX RS485 5 RX RS485 6 RX RS485 7 NC 8 NC 9 NC 10 NC ...

Страница 105: ...Revision E MAC 5500 resting ECG analysis system 5 1 2020299 020 Maintenance 5 Maintenance ...

Страница 106: ...5 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance For your notes ...

Страница 107: ...and possible health hazards The manufacturer does not in any manner assume the responsibility for performing the recommended maintenance schedule unless an Equipment Maintenance Agreement exists The sole responsibility rests with the individuals hospitals or institutions utilizing the device Required Tools and Supplies In addition to a standard set of hand tools you will need the items listed belo...

Страница 108: ...e frequently if needed 1 Use a clean soft cloth and a mild dish washing detergent diluted in water 2 Wring the excess water from the cloth Do not drip water or any liquid on the equipment and avoid contact with open vents plugs or connectors 3 Dry the surfaces with a clean cloth or paper towel Interior Cleaning General Check for dust buildup on the surfaces of the interior circuit boards component...

Страница 109: ...Revision E MAC 5500 resting ECG analysis system 5 5 2020299 020 Maintenance Inspection and Cleaning Thermal Printhead 35A ...

Страница 110: ... button to open the unit 2 Slide the battery release button in the direction of the arrow and lift the battery out 3 Install a new battery and close the unit Patient Cable Replacement 1 Press the internal access button to open the unit 2 Press the connector release tabs and pull the connector loose 3 Pull the cable from the retaining tabs 4 Reassemble the cable by reversing the above steps 36A 37A...

Страница 111: ...ble print out set up for future reference Disconnect the unit from the AC wall outlet and remove the power cord from the unit Remove the battery Remove the chart paper Take strict precautions against electrostatic discharge damage Trolley Disassembly 1 Lock the wheels remove the rear trolley panel then loosen the three captive screws located under the trolley 2 Pull the MAC 5500 up and up toward y...

Страница 112: ...ration Power Supply NOTE A 10 TORX driver is required for disassembly and assembly of the power supply Removal 1 Turn the unit over so the bottom side is up 2 Using a 10 TORX driver remove the three screws holding the power supply in place 3 Lift the power supply to expose the wiring harness and ground wire 4 Remove P2 from J2 on the power supply assembly and the ground wire connection from the po...

Страница 113: ...c and not pinched Top Cover Removal NOTE It is not necessary to remove the Power Supply prior to removing the top cover NOTE A 10 TORX driver is required for disassembly and assembly of the top cover 1 Remove the battery 2 Turn the unit over so the bottom side is up and remove the TORX screw through the hole on the right rear corner of the unit This screw is only visible and accessible with the ba...

Страница 114: ... top cover gently pull the thin strip of plastic free from under the keyboard The entire top assembly is now loose NOTE It may be helpful to rotate the top cover 45 to provide a larger opening to clear the display 9 Carefully lift the top assembly up and clear of the raised display Reassembly 1 Raise the display to the vertical position 2 Make sure the bezel surrounding the rear panel connectors i...

Страница 115: ...ay ground at the left of the hinge rod 6 Slide the display hinge metal rod to the left to release it from the mounting detent A flat blade screw driver may be used to help slide the rod 7 Slightly lift up on the right hand side of the display keyboard assembly and pull the assembly to the right to free the tabs from their mounting slots Do not lift the right side of the display too high or the pla...

Страница 116: ...e right until it locks into the right hinge base 6 Replace the hinge bracket with the two TORX screws removed earlier 7 Replace the screw and display ground at the left of the hinge rod 8 Replace the two TORX mounting screws on the right side of assembly Reassembly 1 Raise the display to the vertical position 2 Make sure the bezel surrounding the rear panel connectors is in place Make sure the rel...

Страница 117: ...e 5 11 4 Disconnect all remaining cable connections to the main PCB These include cables to the following power supply printhead battery connect PCB acquisition module cable 5 Remove the COMM board 6 With a TORX driver remove the mounting screws holding the main PCB in place They are located around the outside edges of the main PCB Set screws aside for mounting new board 7 Remove the harness cable...

Страница 118: ...embly Guidelines 4 Rotate bezel to the upright position as shown below 5 With the new bezel in place replace the top cover by reversing the steps described previously 6 Replace the battery and proceed with software serial number and system setups as described in the following sections ...

Страница 119: ...ss Enter 5 Press Shift F3 The following message is displayed 6 Insert the secure digital card with the application software A message similar to the following displays 7 Press the Enter key A series of messages that describe the update process displays One of three things happens at this point If the boot code is not being updated the following message displays If the boot code is being updated to...

Страница 120: ...he service password and press Enter 4 The Service Only Setup window is displayed 5 Enter the serial number of the system NOTE This is the number which was used when the option codes for this system were generated The number entered here must match the serial number on Programming Primary Boot Programming Over System is Shutting Down Current Boot Version New Boot Version Press Enter to start Instal...

Страница 121: ...rom the System Setup menu 5 Select From SD Card from the Restore Setup menu Restore Options Using the option activation codes for the system restore the options which had been installed on the board which was removed These options are printed on a label located on the bottom of the paper tray NOTE Use the activator codes shown on the label on your system The activator codes shown in the figure bel...

Страница 122: ...option In the rare instance you may need this functionality follow these steps 1 Within the system setup function select Basic System 2 Select Option Activation The Option Activation screen displays 3 In the entry field next to the option to be disabled type x followed by the existing option code The corresponding option will then be disabled To re enable the option remove the x preceding the disa...

Страница 123: ...o the ribbon cable 3 Hold the new printhead FIRMLY in place against the two metal tabs on the printhead mounting plate then tighten the two screws 4 Replace the top cover and power up the unit 5 Go to the Setup menu and enter the new printhead resistance value 6 Run a Writer Test test See Chapter 5 COMM Board Replacement Use the following procedure to replace the communications board 1 Remove the ...

Страница 124: ...le Setting up LAN Communications After replacing the communications board use the following procedure to set up the system s LAN communications NOTE Acquire the IP address gateway and subnet mask from the network administrator before beginning this process 1 Power on the MAC system 2 From the Main Menu select System Setup 3 If prompted enter the system password The default password is system 4 Fro...

Страница 125: ...tely 8 to 10 seconds If it takes less than 8 seconds the settings were not saved successfully Repeat a second time 13 Cycle power on the MAC 3500 MAC 5500 system This ensures the network settings are reread from the communications board 14 Wait at least 10 minutes and verify the MAC 3500 MAC 5500 can communicate with the MUSE system Windows File Server 2000 2003 has an ARP cache that regenerates t...

Страница 126: ...n the paper size bracket to the 8 5 x 11 position 5 Close the unit and turn it over so the bottom side is up 6 Remove the four screws located on the underside of the writer roller carriage assembly and lift the writer from the bottom of the unit Reassembly Reassemble the writer roller carriage assembly by reversing the above procedures Trolley Casters Removal 1 Remove the MAC 5500 and all loose it...

Страница 127: ... an Allen wrench remove the bolt holding the caster to the trolley Reassembly 1 Install the replacement caster on the trolley NOTE Ensure that the pins align with the holes on the fixed caster before fastening to the trolley 2 Install the wheel and attach with the wheel bearing shaft and nut 56A 57A 57A 58A ...

Страница 128: ...em Revision E 2020299 020 Maintenance Disassembly Guidelines 3 Using a small mallet tap the bearing dust covers back in place 4 Set trolley upright and push to check alignment and free movement of casters 5 Replace the MAC 5500 unit 56A 59A ...

Страница 129: ... 5 1 2 7 1 2 3 CAM 14 I F Cable 5 1 2 1 2 3 Battery Pack 1 5 5 1 3 11 Shock Cylinder 1 2 4 5 9 1 3 Motherboard 1 2 3 4 5 3 5 9 10 1 2 3 4 13 20 5 through 9 Roller Assy 1 2 4 5 9 1 2 3 15 Writer Assy 1 2 4 5 9 1 2 3 15 Display Assy 1 2 4 5 3 7 9 1 2 3 12 Keyboard Assy 1 2 4 5 3 9 1 2 3 14 Power Supply Assy 1 2 3 4 5 9 10 1 2 3 Plastics 1 2 3 4 5 7 9 10 1 3 20 Bottom Plastic Replacement Only Trolley...

Страница 130: ... 5 9 1 3 5 LAN Comm PCB 1 2 5 9 1 3 5 19 Barcode Reader 1 2 5 9 1 3 8 Mag Card Reader 1 2 5 9 1 3 9 Treadmill or Ergometer I F Cable 1 2 5 9 1 3 6 Blood Pressure Monitor 1 2 5 9 1 3 Table 4 Option Repairs Continued FRU Option Tools Visual Inspection Checkout Procedure s Table 5 Non FRU Repairs Description Tools Visual Inspection Checkout Procedure s No Parts Replaced 1 5 1 2 3 4 5 6 7 1 3 4 5 thro...

Страница 131: ...l surfaces passed inspection 8 FUSE type and rating passed inspection 9 Fasteners replaced and secured 10 All harnesses and internal wiring has been secured Checkout Procedures Perform the applicable checkout procedure Operational Checks 1 Power up self test passed 2 Simulated recorded rhythm strip successful 3 Simulated recorded ECG successful 4 Simulated ECG stored on media successful 5 Simulate...

Страница 132: ...Battery Current expectation 12 Display diagnostic test successful 13 Speaker test successful 14 Keyboard test successful 15 Writer diagnostic tests successful 16 Read Write Floppy Diagnostic Test successful Accurate test does not have to be performed 17 External Internal Modem Test successful 18 SD Card Test successful 19 Ethernet Module Test successful ...

Страница 133: ...NC ______ Pass Fail 300 Enclosure Leakage Current 1 Forward Polarity NC ______ Pass Fail 100 2 Neutral open Forward Polarity SFC ______ Pass Fail 300 3 Ground open Forward Polarity SFC ______ Pass Fail 300 4 Ground open Reverse Polarity SFC ______ Pass Fail 300 5 Neutral open Reverse Polarity SFC ______ Pass Fail 300 6 Reverse Polarity NC ______ Pass Fail 100 Patient Leakage Current to Ground 1 Fo...

Страница 134: ...ty Resistance 1 AC mains power cord ground prong to exposed metal surface ground lug N A ______ Pass Fail Less than 200m 1 NC Normal Condition SFC Single Fault Condition N A Not Applicable 2 UUT Unit Under Test 3 All SIPs SOPs grounded Table 6 Electrical Safety Checks Step 1 Condition 2 UUT ON Result Leakage Current Limits ...

Страница 135: ...Revision E MAC 5500 resting ECG analysis system 5 31 2020299 020 Maintenance Functional Checkout Procedures For your notes ...

Страница 136: ...5 32 MAC 5500 resting ECG analysis system Revision E 2020299 020 Maintenance Functional Checkout Procedures ...

Страница 137: ...Revision E MAC 5500 resting ECG analysis system 6 1 2020299 020 Parts List 6 Parts List ...

Страница 138: ...6 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List For your notes ...

Страница 139: ...ng Parts General Information The FRU parts lists in this chapter supply enough detail for you to order parts for the assemblies considered field serviceable To order parts contact Service Parts at the address or telephone number on the How to Reach Us page provided at the beginning of this manual ...

Страница 140: ...rence see pn 2026609 001 MAC 5500 Assembly Field Replaceable Units Item Part Number Battery Assembly 900770 001 Power Supply Assembly 421117 001 Keyboard Assembly 421115 XXX Display Assembly 2019106 001 Top Cover See NOTE Label Top Cover See NOTE NOTE Top Cover Label must be ordered together 2017413 001 2008167 001 Printhead 422397 001 Writer Assembly 421108 006 Roller Assembly 422396 006 Writer R...

Страница 141: ...421115 108 Keyboard Assembly Stress Dutch 421115 208 Keyboard Assembly Resting Norwegian 421115 109 Keyboard Assembly Stress Norwegian 421115 209 Keyboard Assembly Resting Danish 421115 110 Keyboard Assembly Stress Danish 421115 210 Keyboard Assembly Resting Czech 421115 111 Keyboard Assembly Stress Czech 421115 211 Keyboard Assembly Resting Hungarian 421115 114 Keyboard Assembly Stress Hungarian ...

Страница 142: ...FT Stress 401855 104 Power Cord Swiss 10A 8FT Stress 401855 107 Power Cord Indian 10A 8FT Stress 401855 108 Power Cord Danish 10A 8FT Stress 401855 109 Power Cord AUST 10A 8FT Stress 401855 110 Power Cord 16A Euro 401855 201 Power Cord 13A British 401855 202 Power Cord 16A Italian 401855 203 Power Cord 16A Israeli 401855 204 Power Cord 15A AUST 401855 210 Cord Power Stress 125V 15A 12FT 405535 002...

Страница 143: ... MAC5000 Spanish 2031240 004 Kit 2D Barcode Imager MAC5000 Swedish 2031240 005 Kit 2D Barcode Imager MAC5000 Italian 2031240 006 Kit 2D Barcode Imager MAC5000 Norwegian 2031240 009 Kit 2D Barcode Imager MAC5000 Danish 2031240 010 Kit 2D Barcode Imager MAC5000 Czech 2031240 011 Kit 2D Barcode Imager MAC5000 Hungarian 2031240 014 Kit 2D Barcode Imager MAC5000 Polish 2031240 015 Barcode Scanner Cable...

Страница 144: ... the CPU board Earlier versions of the software will not work with the 002 board 2022332 002 PCB Comm Board LAN internal modem NOTE This board is not compatible with MAC5500s shipped with the 002 Comm board NOTE Always reload the same or newer version of the software Do not downgrade the software to a lower version 2022332 001 PCB Comm Board LAN internal modem NOTE This board is compatible with MA...

Страница 145: ... government use only 2034530 003 Assembly Wireless Silex Serial Server USA 2034529 001 Assembly Wireless Silex Serial Server Europe 2034529 002 Power Supply Wireless Troy Silex Serial Server for FIPS and non FIPS 2026825 001 Silex antenna for FIPS and non FIPS 2038517 001 Power Supply MAC 5000 Wireless Symbol and Ultra High Security 2013212 001 1 Replaced by and compatible with 2034530 001 Silex S...

Страница 146: ...6 10 MAC 5500 resting ECG analysis system Revision E 2020299 020 Parts List Field Replaceable Units ...

Страница 147: ...Revision E MAC 5500 resting ECG analysis system A 1 2020299 020 Appendix A Abbreviations A Appendix A Abbreviations ...

Страница 148: ...A 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations For your notes ...

Страница 149: ...l aluminum AllRam all RAM AllSec all sector AllTrk all track ALT alternate Alt Off alternate offset am AM acquisition module ante meridiem AM 1 acquisition module 1 AM 1M acquisition module 1 modified AM 2 acquisition module 2 AM 3 acquisition module 3 AM 4 acquisition module 4 amp ampere Ampl amplifier AMU ambulatory monitoring unit ANA analog ANLG analog AnsrTone answer tone A O Analog Output AS...

Страница 150: ...mic CFM cubic feet minute CGR computer graphic record Ch CH channel C L center line CLK clock signal Clonid Clonidine cm centimeter cm2 square centimeters Cmd command number CMMR common mode rejection ratio CMOS complementary metal oxide semiconductor c o in care of COM1 communications port 1 COM2 communications port 2 ComLink communications link Comp composition Confrmd confirmed Cont CONT Contin...

Страница 151: ...double pole single throw DRAM dynamic RAM DR DT digital recording digital transmission DSKTP desktop Dysopyr Dysopyramide E E enable vector electrode site vector lead ecg Ecg ECG electrocardiogram ECO Engineering Change Order EDIC Electrocardiograph Digital Information Center EEPROM electrically erasable programmable ROM e g for example EGA enhanced graphics adapter EMF electromotive force EMI ele...

Страница 152: ...ead HDLC high level data link control Hex HEX hexagon hexadecimal HH hour HiRes high resolution Hr hour Hydral Hydralazine Hz Hertz cycles per second I I on input vector electrode site I II III limb leads IC integrated circuit ID identification i e that is IEC International Electrotechnical Commission in inch IN input inc inc INC incorporated Info information Ins insert I O input output I P input ...

Страница 153: ...mains voltage voltage of a supply mains between 2 line conductors of a polyphase system or voltage between the line conductor and the neutral of a single phase system max maximum Measure measurements Med medications MEM memory MF metal film MHz megahertz min minutes minimum Misc miscellaneous mm millimeter MM minute MMM month mm mV millimeter per millivolt mm s millimeter per second Modem modulato...

Страница 154: ...UT output oz ounce P P P wave section of the ECG waveform p p peak to peak PA P wave amplitude Params parameters Passwds passwords PatData patient data PatInfo patient information PATN patient PC printed circuit personal computer PCB printed circuit board pF picofarad Pgm program PgmId program identification Phenoth Phenothiazide Phenytn Phenytoin PID patient identification digit PLCC plastic lead...

Страница 155: ...eform interval of ventricular depolarization QT QRS interval QTC QRS interval QTY quantity Quinid Quinidine R R resistor red reset RA right angle right arm or R wave amplitude RAM random access memory RC resistor capacitor RD R wave duration Ref reference refresh REN Ringer Equivalence Number Reserp Reserpine REV revision RevdBy reviewed by RevXmit reverse transmission rf radio frequency RFI radio...

Страница 156: ...e end STJ ST segment displacement at the J point STM ST segment displacement at the mid point between STJ and STE stmts Stmts statements supply mains permanently installed power source SVT power cord type 300 V sw SW switch software SW Swedish Sweden T T Tone touch tone TA T wave amplitude Tant tantalum TDML treadmill TE timeout errors Tech technical Thiazid Thiazide TM trademark Tot total TP test...

Страница 157: ...ics array VIA versatile interface adapter VIO violet Volt voltage VRAM video RAM vs versus W w with W watt Warfar Warfarin WHT white WI Wisconsin X x by as in 8 1 2 x 11 XCV transceiver X Y Z orthogonal leads Y Y year yellow yr year yrs years YY year Symbols SHIFTed or alternate function micro F microfarad s sec microsecond 68K 68000 and number C degrees Celsius F degrees Fahrenheit Ohm ohm percen...

Страница 158: ... 5500 resting ECG analysis system Revision E 2020299 020 Appendix A Abbreviations Standard Abbreviations asterisk suffix are active at their relatively higher potential or active high 12SL 12 simultaneous leads ...

Страница 159: ...Revision E MAC 5500 resting ECG analysis system B 1 2020299 020 Appendix B Technical Specifications B Appendix B Technical Specifications ...

Страница 160: ...B 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications For your notes ...

Страница 161: ...ty to archive ECG records on removable media SD card Digital Sampling Rate 4000 samples s channel Analysis Pediatric and vectorcardiography Optional 12SL analysis HI RES and PHI RES late potential analysis Pre acquisition Provides 10 s of instantaneous ECG acquisition Dynamic Range AC differential 5 mV DC offset 320 mV Resolution 4 88 µV LSB 250 sps 1 22 µV LSB 500 sps Frequency Response 3 dB 0 01...

Страница 162: ...board Item Description Type Sealed elastomer with soft function keys alphanumeric keys writer controls and TrimPad cursor controls Electrical Item Description Power Supply AC or battery operation AC Input Voltage Current Frequency 100 240 VAC 10 15 0 5 A 115 VAC 0 3 A 240 VAC typical 50 60 Hz 10 Battery Type User replaceable 18V 3 5 AH 10 rechargeable NiMH pack Battery Capacity 100 single page rep...

Страница 163: ... 20 mm mV and 50 mm mV 1 mm µV Analysis Sampling Rate 1000 samples per second per channel Digital Sampling Rate 4000 samples per second per channel High Low Pass Filters Spectral filter using Fast Fourier Transform FFT ADC Resolution 1 22 µV LSB Analysis Resolution 0 1525 µV LSB Physical 1 Item Description Height 9 4 cm 3 7 in with display closed Width 38 1 cm 15 0 in Depth 35 1 cm 13 8 in Weight ...

Страница 164: ...o 95 RH noncondensing 500 to 1060 hPa Disposal Batteries Device Disposing of battery by fire or burning will cause the battery to explode The battery is recyclable Follow local environmental guidelines concerning disposal and recycling Batteries may be returned to GE Medical Systems Information Technologies service for recycling Recyclable 1 Paper discoloration may occur at higher temperatures Env...

Страница 165: ...zardous materials Patient Mode of Operation Continuous Patient Leakage Current 10 µA Degree of Protection Against Electrical Shock Type BF defibrillation protection for the patient cable acquisition module Maintenance Frequency Daily visual inspection and routine cleaning if needed performed by user Use a commercially available industrial strength disinfectant cleaner on any part of the equipment ...

Страница 166: ...B 8 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix B Technical Specifications Technical Specifications For your notes ...

Страница 167: ...Revision E MAC 5500 resting ECG analysis system C 1 2020299 020 Appendix C Electromagnetic Compatibility C Appendix C Electromagnetic Compatibility ...

Страница 168: ...C 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility For your notes ...

Страница 169: ...uipment or system should be tested to verify normal operation in the configuration in which it is being used Guidance and Manufacturer s Declaration Electromagnetic Emissions The MAC 5500 is intended for use in the electromagnetic environment specified below It is the responsibility of the customer or user to ensure that the MAC 5500 is used in such an environment Emissions Test Compliance Electro...

Страница 170: ...y lines 1 kV for input output lines Mains power should be that of a typical commercial or hospital environment Surge EN 61000 4 5 1 kV differential mode 2 kV common mode 1 kV differential mode 2 kV common mode Mains power should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines EN 61000 4 11 5 Ut 95 dip in Ut...

Страница 171: ... is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site surveya should be less than the compliance level in each frequency rangeb Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in ...

Страница 172: ...ipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts w according to the transmitter manufacturer NOTE These guidelines may not apply in all situations Electromagnetic propagation ...

Страница 173: ... Host Mac 5500ST 5 4m 901142 001 Kit Cam 14 Resting W AHA Adaptors NA 901142 002 Kit Cam 14 Resting W IEC Adaptors NA 400073 001 Serial Comm Cable 8 Pin Mini DIN 6 1m 416070 001 External Video Cable 1 8m 700520 002 Analog TTL Interface Cable 3 0m 2007918 001 Treadmill Cable Mac 5500ST to T2000 6 0m 405535 006 Power Supply Cord US 13A 125V 3 0m 401855 001 Power Supply Cord European 10A 250V 2 5m 40...

Страница 174: ...C 8 MAC 5500 resting ECG analysis system Revision E 2020299 020 Appendix C Electromagnetic Compatibility Electromagnetic Compatibility EMC For your notes ...

Страница 175: ...Revision E MAC 5500 resting ECG analysis system Index 1 2020299 020 Index ...

Страница 176: ...Index 2 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...

Страница 177: ... Disk 4 20 G green AC power light 2 4 H how to format a secure data card 4 20 I internal access button 2 4 K keyboard 2 3 L LAN port 2 3 list of abbreviations A 3 location number 4 21 M Modem port 2 3 modem test 4 15 O operating conditions B 5 P pace detection B 3 paper B 4 passwords 4 22 patient data 4 23 patient questions 4 21 product code 1 9 S screen 2 3 Secure data card formatting 4 20 Secure...

Страница 178: ...Index Index 4 MAC 5500 resting ECG analysis system Revision E 2020299 020 ...

Страница 179: ......

Страница 180: ...ical Systems Information Technologies GmbH Munzinger Straße 3 5 D 79111 Freiburg Germany Tel 49 761 45 43 0 Fax 49 761 45 43 233 Asia Headquarters GE Medical Systems Information Technologies Asia GE China Co Ltd No 1 Huatuo Road Zhangjiang Hi Tech Park Pudong Shanghai People s Republic of China 201203 Tel 86 21 3877 7888 Fax 86 21 3877 7402 www gehealthcare com ...

Отзывы: