2-6
Section 2-2 - General Console Requirements
2-2-2-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2-3
Time and Manpower Requirements
Site preparation takes time. Begin Pre-installation checks as soon as possible, if possible, six weeks
before delivery, to allow enough time to make any changes.
Table 2-11 LOGIQworks Optional Hardware
PART
Modem
BTU/h
164
Power (W)
48
Total BTU/h
All
x
164
Table 2-12 Average Installation Time
Description
Average Installation
Time
Comments
Unpacking
15 minutes
1 person
Checking delivery
15 minutes
1 person
Installation of
LOGIQworks
1.0 hour
1 person
Check connectivity
and setup
15 minutes
1 person