Chapter 4 Functional Checks
4-7
4-3-1-2
Check Ethernet Connection
Check Ethernet connection (only needed if DCM Ping - see
Section 4-3-1-1 "Check the DICOM
- fails).
Condition: The LOGIQworks system needs to be connected to the network. See
for details.
Check the status of the Ethernet connection to the devices your LOGIQworks system is set up to
communicate with. Please look up all sending and receiving devices in the connections destination and
origins list from the data selector menu. Refer to
and
for detailed information on how to set up sending and receiving devices in LOGIQworks.
Run the network ping command to perform this verification:
1.) Click on
Start
>
Programs
>
Accessories
>
Command Prompt
. A command window will
open.
2.) Execute the ping command for every IP address listed in the destinations and origins database
and
).
Example:
Figure 4-5 Ethernet Connection Check
successful,
Ethernet connection
is good
failed,
connection not
established