GE M
EDICAL
S
YSTEMS
D
IRECTION
5184024-100, R
EVISION
5
LOGIQ
WORKS
® B
ASIC
S
ERVICE
M
ANUAL
Chapter 2 Pre-Installation
2-7
2-2-4
Electrical Requirements
2-2-4-1
LOGIQworks Power Requirements
Electrical Specifications for LOGIQworks
Table 2-13 Electrical Specifications of LOGIQworks System based on Dell Precision 340
2-2-4-2
Inrush Current
230 VAC:2.5A
100 VAC: 5A
2-2-4-3
Site Circuit Breaker
It is recommended that the branch circuit breaker for the machine be readily accessible.
2-2-4-4
Site Power Outlets
Three to five AC power outlet must be less than 1.2 meters (3.9feet) from the LOGIQworks Workstation.
Electrical installation must meet all current local, state, and national electrical codes.
NOTE
LOGIQworks requires a dedicated power and ground for the proper operation of its workstation. This
dedicated power shall originate at the last distribution panel before the system. The dedicated line shall
consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the
distribution panel to the workstation. Please note that image artifacts can occur, if at any time within the
facility, the ground from the main facility's incoming power source to the workstation is only a conduit.
GEVU P/N
Voltage
Tolerances
Power
Frequency
H49021LA
110 - 230 VAC
±10%
398W
50-60 Hz
H49021LB
110 - 230 VAC
±10%
543W
50-60 Hz
H49021LC
110 - 230VAC
±10%
310W
50-60 Hz
H49021LD
110 - 230VAC
±10%
367W
50-60 Hz
Table 2-14 Electrical Specifications for LOGIQworks System based on Dell Precision 360 or
on Dell Precision 370
GEVU P/N
Voltage
Tolerances
Power
Frequency
H49021AA/AC
110 - 230VAC
±10%
310W
50-60 Hz
H49021AB/AD
110 - 230VAC
±10%
367W
50-60 Hz
Table 2-15 Electrical Specifications for LOGIQworks System based on Dell Precision 380 or
on Dell Precision 390
GEVU P/N
Voltage
Tolerances
Power
Frequency
H49021AK/AM
110 - 230VAC
±10%
375W
50-60 Hz
H49021AL/AN
110 - 230VAC
±10%
375W
50-60 Hz
Содержание LOGIQworks
Страница 2: ......
Страница 9: ...GE HEALTHCARE DIRECTION 5184024 100 REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL vii...
Страница 418: ...GE MEDICAL SYSTEMS DIRECTION 5184024 100 REVISION 5 LOGIQWORKS BASIC SERVICE MANUAL II...
Страница 419: ......