Performing an Exam
2-8
LOGIQ E9
–
User Guide
Direction 5454884-100 English
Rev. 1
Doppler Mode Controls
Table 2-4:
Doppler Mode Controls
Control
Possible
Bioeffect
Description/
Benefit
Update
Yes
Toggles between simultaneous and update presentation while
viewing the timeline.
SV Length
(Trackball)
Yes
Moves the sample volume gate on the B-Mode's Doppler Mode
cursor. The gate is positioned over a specific position within the
vessel.
Doppler sample
volume length (SV
Length)
Yes
Sizes the sample volume gate.
Scale (Velocity
Scale)
Yes
Adjusts the velocity scale to accommodate faster/slower blood flow
velocities. Velocity scale determines pulse repetition frequency.
If the sample volume gate range exceeds single gate Scale
capability, the system automatically switches to high PRF mode.
Multiple gates appear, and HPRF is indicated on the display.
Angle Correct/Auto
Angle
No
Estimates the flow velocity in a direction at an angle to the Doppler
vector by computing the angle between the Doppler vector and the
flow to be measured.
NOTE: When the Doppler Mode Cursor and angle correct indicator
are aligned (the angle is O), you cannot see the angle correct
indicator.
Quick Angle
No
Quickly adjusts the angle by 60 degrees.
Wall Filter
No
Insulates the Doppler signal from excessive noise caused from
vessel movement.
Baseline
No
Adjusts the baseline to accommodate faster or slower blood flows to
eliminate aliasing.
Mode Cursor
No
Displays the Doppler Mode cursor on the B-Mode image.
Steer and Fine Steer
Yes
You can slant the ROI of the Color Flow linear image left or right to
get more information without moving the probe. The angle steer
function only applies to linear probes.
Audio Volume
No
Controls audio output.
Invert
No
Vertically inverts the spectral trace without affecting the baseline
position.
Compression
(Dynamic Range)
No
Dynamic range controls how echo intensities are converted to
shades of gray, thereby increasing the range of contrast you can
adjust.
Trace Method
(Spectral Trace)
No
Traces the average mean and peak velocities in realtime or frozen
images.
Cycles to Average
No
The average value over a number of cycles (from 1-5).
Trace Sensitivity
No
Adjust the trace to follow the waveform for signal strength.
Содержание LOGIQ E9
Страница 4: ...i 2 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 This page intentionally left blank ...
Страница 8: ...i 6 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 12: ...i 10 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 126: ...Performing an Exam 2 40 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 202: ...After the Exam is Over 3 76 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 250: ...Safety 4 48 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 254: ...Index 4 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 255: ......
Страница 256: ......