The Biopsy Procedure
1. Place coupling gel on the scanning surface of the probe/
sheath/biopsy guide assembly.
2. Activate the biopsy guidezone on the system through the
B-Mode Touch Panel. When using multi-angle guides,
ensure that the proper guidezone angle is displayed.
Figure 1-43.
B-Mode Touch Panel
3. Scan to locate the target. Center the target in the electronic
guidezone path.
NOTE:
Enabling color flow would allow for visualization of the
vascular structure around the area to be biopsied.
4. Place the needle in the guide between the needle barrel and
needle clip. Direct it into the area of interest for specimen
retrieval.
WARNING
Biopsy procedures must only be performed on live images.