Probes
LOGIQ E9
–
User Guide
1-61
Direction 5454884-100 English
Rev. 1
4D Probe Biopsy Needle Path Selection
To select the needle path and verify that the path of the needle is
accurately indicated within the guidezone on the system
monitor, perform the following before use:
1. Properly install the bracket and biopsy guide.
2. Scan in a container filler with water (47° C).
3. Select
Biopsy kit
. The available biopsy options from the
Touch Panel.
Select the biopsy guidezone where the needle echo passes
through the center of the guidezone. Use the selected
biopsy guidezone when performing the biopsy.
Biopsy Needle Path Verification
To verify that the path of the needle is accurately indicated within
the guidezone on the system monitor, perform the following:
•
Properly install the bracket and biopsy guide.
•
Scan in a container filled with water (47° C).
•
Display the biopsy guidezone on the monitor.
•
Ensure that the needle echo falls within the guidezone
markers.
Содержание LOGIQ E9
Страница 4: ...i 2 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 This page intentionally left blank ...
Страница 8: ...i 6 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 12: ...i 10 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 126: ...Performing an Exam 2 40 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 202: ...After the Exam is Over 3 76 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 250: ...Safety 4 48 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 254: ...Index 4 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 255: ......
Страница 256: ......