Probes
LOGIQ E9
–
User Guide
1-59
Direction 5454884-100 English
Rev. 1
Endocavitary Probe Biopsy Guide Preparation
1. If a biopsy is to be performed, snap the metal or plastic
biopsy guide on to the probe over the sheath.
Figure 1-40.
Disposable Biopsy Guide 5 degree Angle
Figure 1-41.
Reusable Biopsy Guide
a. Fix with a screw
NOTE:
For the RIC5-9-D and IC5-9-D probes, use the TR5
guidelines for the plastic (disposable-only) biopsy guides;
use the RU guidelines with the stainless steel reusable
biopsy guides.
2. Place an adequate amount of ultrasound gel on the gel-filled
sheath tip’s outer surface.
3. Ensure the guide is properly seated and secure by pushing
forward on the needle insertion end of the guide until the
attachment node is firmly in place in it’s hole.
CAUTION
Patient injury or repeated biopsies may result. The needle
placement will not be as intended if the needle guide is not
properly seated and secure.
a
Содержание LOGIQ E9
Страница 4: ...i 2 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 This page intentionally left blank ...
Страница 8: ...i 6 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 12: ...i 10 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 126: ...Performing an Exam 2 40 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 202: ...After the Exam is Over 3 76 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 250: ...Safety 4 48 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 254: ...Index 4 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 255: ......
Страница 256: ......