System Start-Up
LOGIQ E9
–
User Guide
1-25
Direction 5454884-100 English
Rev. 1
Power Off
To power off the system:
1. Set the brake and use the operator panel movement
controls to lock the control panel in place.
2. When you shutdown the system, enter the scan screen and
lightly press the
Power On/Off
switch at the front of the
system once. The System-Exit window is displayed.
NOTE:
DO NOT press and hold down the Power On/Off switch to
shutdown the system. Instead, lightly press the Power On/
Off switch and select Shutdown.
3. Using
the
Trackball
, select Shutdown.
The shutdown process may take up to two (2) minutes and
is completed when the control panel illumination shuts
down.
NOTE:
If the system appears non-responsive, press the Power On/
Off switch. After about six (6) seconds, the “Application not
responding” system dialog appears. Select to end all
system processes and shut down the system.
4. Disconnect the probes.
Clean or disinfect all probes as necessary. Store them in
their shipping cases or another appropriate probe storage
system to avoid damage.
NOTE:
When the Extended Power Supply (EPS) is installed, battery
replacement may be necessary if the circuit breaker is turned off
for long periods of time (3 to 6 months).
Additional System Information
For optimum system operation, we recommend that you restart
the system at least once every 24-hour period. If you shut down
the system at the end of the day, no other action is needed.
HINTS
If the Park Lock does not engage when you attempt to
dock the Operator Panel in place, reset the Park Brake by
selecting “Park Brake Reset” in the Miscellaneous section
on the Utility -> System -> Peripherals page.
CAUTION
DO NOT
turn off the circuit breaker before the Power On/Off
switch LED is off.
Data may be lost or system software damaged if the circuit
breaker is turned off before the Power On LED is off.
Содержание LOGIQ E9
Страница 4: ...i 2 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 This page intentionally left blank ...
Страница 8: ...i 6 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 12: ...i 10 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 126: ...Performing an Exam 2 40 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 202: ...After the Exam is Over 3 76 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 250: ...Safety 4 48 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 254: ...Index 4 LOGIQ E9 User Guide Direction 5454884 100 English Rev 1 ...
Страница 255: ......
Страница 256: ......