Testing instructions for KISS™ and KISS™ Multilead
The electrodes must adhere securely.
4.
Changing the pressure with the regulator results
in a visible effect
on the suction cups.
Negative pressure regulator in position
(left), the electrodes should drop off.
Equipment Check for Leaks in the Complete
System, Pump, Distributor, Electrode Leads and
Suction Cups
Testing Requirements
• Pneumatic Transducer Tester PTT, e.g., X-Caliber (shown in mmHg)
• Ohmmeter/Multimeter
• Test plate for connection of a suction electrode
see
“1. Test Plate (p/n 30344490)” on page 45
• Y-adapter (for connection negative pressure tubing <-> PTT)
see
“2. Y-Adapter (p/n 30344489)” on page 45
• Dummy plug for distributor
see
“4. Dummy Plug for KISS™ (p/n 30344488)” on page 46
20
KISS™ / KISS™ Multilead
22745010H