Testing instructions for KISS™ and KISS™ Multilead
Remedy: Replace electrode; see
See Chapter 5. “Spare Parts List (KISS™/KISS™
Multilead)” on page 41.
Testing Electrical Function (KISS
™
™
™
)
Resistance between the leadwire connector and the KISS™ electrode connector. The
measured resistance must be 10 kohm ±1kohm.
CAUTION:
It is prudent to make this measurement without electrodes as the silver-silver
chloride electrodes must not come into contact with other metals.
Pin assigment for ECG units with IEC plug:
R=9, L=10, F=11, N=14, C1=12, C2=1, C3=2, C4=3, C5=4, C6=5, shield=6
24
KISS™ / KISS™ Multilead
22745010H